Related Experiment Video
Updated: Jun 7, 2026

Evaluating Targeting Accuracy in the Focal Plane for an Ultrasound-guided High-intensity Focused Ultrasound Phased-array System
Published on: March 6, 2019
A pre-clinical phantom comparison of tissue harmonic and brightness mode imaging for application in ultrasound guided
G K Sandhu1, P B Dunscombe, R F H Khan
1Department of Medical Physics, Tom Baker Cancer Centre, Calgary, Alberta T2N 4N2, Canada. Gurpreet.Sandhu2@albertahealthservices.ca
Purpose:
The current practice of prostate brachytherapy utilizes the brightness (B) mode ultrasound imaging for volume definition and needle guidance. However, tissue harmonic (H) mode available with new scanners has shown the improved image quality. The aim of this study was to perform a pre-clinical phantom evaluation of harmonic imaging as an alternative to B mode in prostate brachytherapy.
Methods:
Performance characteristics viz. dead zone, depth of penetration, geometric accuracy, spatial resolution, tissue to clutter ratio (TCR) and signal to noise ratio (SNR), were compared between two modes using an in-house phantom. Images were acquired under the same settings except the gain; which is higher for the H mode than that of B mode. A qualitative comparison between two modes was also performed using commercial CIRS053 phantom.
Results:
Dead zone, depth of penetration and geometric accuracy were respectively <1 mm, >8 cm and <1% for both modes. Relative TCR, SNR and the spatial resolution were improved in H mode compared with B mode. Images with CIRS053 phantom in H mode demonstrate sharper boundaries for prostate and urethra, freedom from background clutter, and better identification of the brachytherapy needles.
Conclusions:
This study indicates the superiority of H over B mode, in terms of spatial resolution, relative contrast, and overall image quality. Thus H mode has the potential benefit in prostate brachytherapy. This study provides the basis to move forward to investigate whether the superior image quality observed in the laboratory can be translated into a higher treatment quality for the patient.
More Related Videos
06:33Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
06:08A Cognitive Fusion-guided Prostate Biopsy Using Multiparametric Magnetic Resonance Imaging and Transrectal Ultrasound
Published on: March 21, 2025