Echocardiographic Assessment Using Subxiphoid-Only Examination for Hypotensive Patients
Published on: April 18, 2025
New pocket echocardiography device is interchangeable with high-end portable system when performed by experienced
C A Frederiksen1, P Juhl-Olsen, U T Larsen
1Department of Anaesthesiology and intensive care, Aarhus University Hospital, Skejby, Denmark.
Background:
Cardiovascular status is a crucial determinant in the pre-operative assessment of patients for surgery as well as for the handling of patients with acute illness. We hypothesized that focus-assessed transthoracic echocardiography (FATE) could be performed with the subject in the semi-recumbent position. The aim was also to test whether the image quality of Vscan is interchangeable with a conventional high-quality portable echocardiography system. Furthermore, we evaluated the time needed to achieve an interpretable four-chamber view and to complete a full FATE examination.
Methods:
Sixty-one subjects were included. All subjects were examined in accordance with the FATE protocol in the semi-recumbent position on two different systems: the novel Vscan pocket device and the high-quality portable Vivid i system. Two evaluations were performed. In group A (n=30), the focus was on image quality. In group B (n=31), the focus was on the time consumed.
Results:
Group A: All patients (100%) had at least one image suitable for interpretation and no significant difference in image quality (P=0.32) was found between the two different systems. Group B: The mean value for the total time consumed for a full FATE was 69.3 s (59.8-78.8) on the Vscan and 63.7s (56.7-70.8) on the Vivid i, with no significant difference among the scanners (P=0.08).
Conclusion:
The Vscan displays image quality interchangeable with larger and more expensive systems. The apparatus is well suited for performing a FATE examination in a 1-day surgery setting and could very well also be applicable in almost any situation involving patients with acute illness.
Related Concept Videos
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion, evaluates...
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for diagnosing...
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...