A comparison between Monte Carlo-calculated and -measured total efficiencies and energy resolution for large plastic scintillators used in whole-body counting
Related Experiment Video
Updated: Jun 7, 2026

Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera
Published on: January 30, 2020
1Department of Radiation Physics, Institute of Clinical Sciences, The Sahlgrenska Academy, University of Gothenburg, SU/Sahlgrenska, 413 45 Gothenburg, Sweden.
Abstract:
The measured total efficiency for several source positions has been determined for a large, plastic scintillation detector (NE 102A, 91.5 × 76.0 × 24.5 cm(3)) used for whole-body counting gamma spectrometry. The results have been compared with Monte Carlo-calculated total efficiencies; the code used was MCNPX 2.60. For (137)Cs, there was a good agreement between the measured and calculated total efficiencies. MXNPX was also used to calculate the electron light yield for (137)Cs; for the detector material NE 102A, Birks' constant kB was found to be 9.6 mg cm(-2) MeV(-1). The effect of light losses on spectrum resolution has been investigated for (65)Zn.