Removal of ring artifacts in CT imaging through detection and correction of stripes in the sinogram
Emran Mohammad Abu Anas1, Soo Yeol Lee, Md Kamrul Hasan
1Department of Electrical and Electronic Engineering, Bangladesh University of Engineering and Technology, Dhaka-1000, Bangladesh.
Abstract:
Due to malfunctioning and mis-calibration of cells in digital x-ray detectors as well as impurities on the scintillator screens, stripe artifacts arise in the sinogram which in turn generate ring artifacts in the reconstructed x-ray computed tomography images. In this paper, a novel technique is proposed for the detection and removal of stripe artifacts in a sinogram with a view to suppress the ring artifacts from the tomographic images. To accurately detect the stripe creating pixels using a derivative-based algorithm, at first the sinogram is windowed to create a sub-sinogram by keeping the pixel of examination at the center position in the sub-sinogram. The other pixels in the sub-sinogram are selected from a polyphase component of the sinogram. A new mathematical index is proposed here to isolate the strong and weak ring-generating stripes from the good ones. For the correction of strong ring artifacts resulting from the defective detector elements and dusty scintillator crystals, 2D variable window moving average and weighted moving average filters are proposed in this work. On the other hand, a conventionally trusted constant bias correction scheme is adopted to correct the responses of the mis-calibrated detector elements. To evaluate and compare the performance of the proposed algorithm, real micro-CT images acquired from two flat panel detectors under different operating conditions are used. Experimental results show that the proposed method can remove ring artifacts more effectively without imparting noticeable distortion in the image as compared to a recently reported technique in the literature.
More Related Videos
10:23Equipment Setup and Artifact Removal for Simultaneous Electroencephalogram and Functional Magnetic Resonance Imaging for Clinical Review in Epilepsy
Published on: June 23, 2023
07:13Multimodal Cross-Device and Marker-Free Co-Registration of Preclinical Imaging Modalities
Published on: October 27, 2023
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...