Published on: December 18, 2016
Short-term repeatability of joint space width measurements using a magnetic resonance imaging compatible knee
M F Koff1, D W Stanley, P J Weishaar
1Department of Radiology and Imaging, Hospital for Special Surgery, USA.
Summary
An MRI-compatible knee positioning device improved joint space width measurement consistency between technologists for the tibiofemoral joint. This device may enhance accuracy in knee joint health studies.
Area of Science:
- Radiology
- Orthopedics
- Medical Imaging
Background:
- Accurate joint space width (JSW) measurements are crucial for assessing knee joint health.
- Variability in JSW measurements can arise from differences in technologist positioning during MRI scans.
- Standardized positioning may improve the reliability of quantitative MRI assessments.
Purpose of the Study:
- To evaluate an MRI-compatible knee positioning device.
- To assess the device's ability to minimize intratechnologist and intertechnologist differences in minimum JSW measurements.
- To determine the impact of the device on scan-to-scan, intratechnologist, and intertechnologist repeatability.
Main Methods:
- Five subjects underwent MRI scans with and without an MRI-compatible positioning device.
- Two technologists performed scans, with repositioning between some scans.
- A semi-automated program calculated minimum JSW for tibiofemoral and patellofemoral joints.
- Scan-to-scan, intratechnologist, and intertechnologist repeatability were assessed.
- Root mean square (RMS) error was calculated for JSW measurements.
Main Results:
- Scan-to-scan and intratechnologist repeatability remained unchanged with the device.
- Intertechnologist repeatability for the tibiofemoral joint improved (0.70 to 0.42 mm).
- RMS error for the tibiofemoral joint significantly decreased (0.26 to 0.15 mm, P = 0.0006).
- Patellofemoral JSW measurement variability slightly increased but was not statistically significant.
Conclusions:
- The MRI-compatible positioning device effectively reduced JSW measurement variability for the tibiofemoral joint.
- Increased variability in patellofemoral measurements may relate to articular surface incongruities.
- The device shows potential benefit for quantitative, longitudinal knee joint health studies.