Related Experiment Video
Updated: Jun 6, 2026

Correction of Presbyopia by Monocular Bi-Aspheric Ablation Profile
Published on: September 20, 2024
Coma correction and extension of the focusing geometry of a soft-x-ray monochromator
Abstract:
I show analytically and numerically how to use a varied line-space grating to reduce coma aberration in a soft-x-ray synchrotron radiation monochromator: a plane grating in combination with one or more spherical mirrors arranged in front of the grating and rotated around a fixed axis outside the mirror surface. It seems that the variation extent of groove spacing is too small to be manufactured; thus, a holographic grating of equivalent imaging properties is proposed. Based on the analysis of a focusing equation, I found that, if the plane grating is replaced by a spherical one, the monochromator can also be made free of defocus.
More Related Videos
07:24Quantitative Atomic-Site Analysis of Functional Dopants/Point Defects in Crystalline Materials by Electron-Channeling-Enhanced Microanalysis
Published on: May 10, 2021
11:27Studying Soft-matter and Biological Systems over a Wide Length-scale from Nanometer and Micrometer Sizes at the Small-angle Neutron Diffractometer KWS-2
Published on: December 8, 2016