Vapor-phase hydrogen peroxide as a surface decontaminant and sterilant
1School of Public Health, University of Minnesota, Minneapolis 55455.
Abstract:
The feasibility of utilizing vapor-phase hydrogen peroxide (VPHP) as a surface decontaminant and sterilant was evaluated in a centrifuge application. The prototype VPHP decontamination system, retrofitted into a Beckman L8-M ultracentrifuge, was designed to vaporize a 30% (wt/wt) solution of aqueous hydrogen peroxide continuously injecting and withdrawing VPHP in a deep-vacuum flow-through system. VPHP cycles of 4, 8, 16, and 32 min were examined for cidal activity against spores of Bacillus subtilis subsp. globigii and Bacillus stearothermophilus. Spore inocula (approximately 10(6)/coupon) were dried onto 0.5-in. (1.27-cm)-square stainless-steel coupons, and coupons were suspended in the centrifuge chamber, the space between the refrigeration can and the barrier ring (inner gap), and the space between the barrier ring and the vacuum ring (outer gap). At a chamber temperature of 4 degrees C, B. subtilis subsp. globigii spores were inactivated within 8 min, while inactivation of spores located in the outer gap at 27 degrees C required 32 min. The elevated temperature and high surface area/volume ratios in the outer gap may serve to decompose the gas more rapidly, thus reducing cidal efficacy. Of the two test spores, B. stearothermophilus was more resistant to VPHP. Nonetheless, VPHP was shown to possess significant sporicidal capability. For practical decontamination applications of the type described, VPHP shows promise as an effective and safer alternative to currently used ethylene oxide or formaldehyde vapors.
Related Concept Videos
Phase Transitions: Vaporization and Condensation
Autoxidation of Ethers to Peroxides and Hydroperoxides
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Chemical Agents for Microbial Control