Automated surveillance to detect postprocedure safety signals of approved cardiovascular devices
Frederic S Resnic1, Thomas P Gross, Danica Marinac-Dabic
1BrighamandWomen’s Hospital and Harvard Medical School, Boston, Massachusetts 02115, USA. fresnic@partners.org
Context:
Ensuring the safety of medical devices challenges current surveillance approaches, which rely heavily on voluntary reporting of adverse events. Automated surveillance of clinical registries may provide early warnings in the postmarket evaluation of medical device safety.
Objective:
To determine whether automated safety surveillance of clinical registries using a computerized tool can provide early warnings regarding the safety of new cardiovascular devices.
Design, Setting, And Patients:
Prospective propensity-matched cohort analysis of 7 newly introduced cardiovascular devices, using clinical data captured in the Massachusetts implementation of the National Cardiovascular Data Repository CathPCI Registry for all adult patients undergoing percutaneous coronary intervention from April 2003 through September 2007 in Massachusetts.
Main Outcome Measure:
Presence of any safety alert, triggered if the cumulative observed risk for a given device exceeded the upper 95% confidence interval (CI) of comparator control device. Predefined sensitivity analyses assessed robustness of alerts when triggered.
Results:
We evaluated 74,427 consecutive interventional coronary procedures. Three of 21 safety analyses triggered sustained alerts in 2 implantable devices. Patients receiving Taxus Express2 drug-eluting stents experienced a 1.28-fold increased risk of postprocedural myocardial infarction (2.87% vs 2.25%; absolute risk increase, 0.62% [95% CI, 0.25%-0.99%]) and a 1.21-fold increased risk of major adverse cardiac events (4.24% vs 3.50%; absolute increase, 0.74% [95% CI, 0.29%-1.19%]) compared with those receiving alternative drug-eluting stents. Patients receiving Angio-Seal STS vascular closure devices experienced a 1.51-fold increased risk of major vascular complications (1.09% vs 0.72%; absolute increased risk, 0.37% [95% CI, 0.03%-0.71%]) compared with those receiving alternative vascular closure devices. Sensitivity analyses confirmed increased risk following use of the Taxus Express2 stent but not the Angio-Seal STS device.
Conclusion:
Automated prospective surveillance of clinical registries is feasible and can identify low-frequency safety signals for new cardiovascular devices.
More Related Videos
16:40A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
06:10Utilizing Percutaneous Ventricular Assist Devices in Acute Myocardial Infarction Complicated by Cardiogenic Shock
Published on: June 12, 2021
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Pharmacovigilance
This process, termed pharmacovigilance, aims to detect, evaluate, and minimize harmful effects related to medication use. The data collection for pharmacovigilance depends on spontaneous reporting systems, where healthcare professionals or patients voluntarily report suspected ADRs.
In some cases, there...
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion,...
Imaging Studies for Cardiovascular System V: CT
Holter Monitor: 24-Hour Monitoring
Cardiac Catheterization I: Pre-Procedure Overview