Jove
Visualize
Contact Us
JoVE
x logofacebook logolinkedin logoyoutube logo
ABOUT JoVE
OverviewLeadershipBlogJoVE Help Center
AUTHORS
Publishing ProcessEditorial BoardScope & PoliciesPeer ReviewFAQSubmit
LIBRARIANS
TestimonialsSubscriptionsAccessResourcesLibrary Advisory BoardFAQ
RESEARCH
JoVE JournalMethods CollectionsJoVE Encyclopedia of ExperimentsArchive
EDUCATION
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab ManualFaculty Resource CenterFaculty Site
Terms & Conditions of Use
Privacy Policy
Policies

Related Concept Videos

Pipe Flowrate Measurement01:28

Pipe Flowrate Measurement

In pipe flow measurement, orifice, nozzle, and Venturi meters are commonly used to determine fluid flowrates by constricting the flow area, which increases fluid velocity and reduces pressure. This pressure difference, governed by Bernoulli's principle and adjusted for real-world conditions, is essential for calculating flowrate. Each meter type is suited to specific applications based on accuracy, efficiency, and compatibility with various flow conditions.
The orifice meter is a simple,...
Bernoulli's Principle: Applications01:17

Bernoulli's Principle: Applications

There are many devices and situations in which fluid flows at a constant height and so can be analyzed using Bernoulli's principle. These devices include, but are not limited to, entrainment devices and fluid flow measuring devices.
Entrainment devices use a high fluid speed to create low pressures and, thus, entrain one fluid into another. Some examples of these devices are given below:
Pressure Gauges01:20

Pressure Gauges

Most pressure gauges, like those on scuba tanks, are calibrated to read zero at atmospheric pressure. Readings from such gauges are called the gauge pressure, which is the pressure relative to atmospheric pressure. When the pressure inside the tank exceeds atmospheric pressure, the gauge reports a positive value. Some gauges are designed to measure negative pressure. For example, many physics experiments must take place in a vacuum chamber, a rigid chamber from which some of the air is pumped...

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Perinatal factors influencing the earliest establishment of the infant microbiome.

Microbiome research reports·2025
Same author

Late-onset sepsis treatment in very preterm infants alters longitudinal microbiome trajectory with lower abundance of <i>Bifidobacterium</i> despite probiotic supplementation.

Gut microbes·2025
Same author

Observational study protocol: the faecal microbiome in the acute stage of new-onset paediatric type 1 diabetes in an Irish cohort.

BMJ open·2025
Same author

Somatic cell count as an indicator of subclinical mastitis and increased inflammatory response in asymptomatic lactating women.

Microbiology spectrum·2024
Same author

Delaying cord clamping for more than 1 minute after vaginal deliveries appeared to result in smoother cardiocirculatory transition.

Acta paediatrica (Oslo, Norway : 1992)·2024
Same author

First encounters of the microbial kind: perinatal factors direct infant gut microbiome establishment.

Microbiome research reports·2023

Related Experiment Video

Updated: Jun 6, 2026

Protocol for Relative Hydrodynamic Assessment of Tri-leaflet Polymer Valves
11:12

Protocol for Relative Hydrodynamic Assessment of Tri-leaflet Polymer Valves

Published on: October 17, 2013

The Neopuff's PEEP valve is flow sensitive.

Colin Patrick Hawkes1, Eugene Michael Dempsey, C Anthony Ryan

  • 1Department of Neonatology, Cork University Maternity Hospital, Cork, Ireland.

Acta Paediatrica (Oslo, Norway : 1992)
|November 13, 2010
PubMed
Summary

The Neopuff's positive end-expiratory pressure (PEEP) valve is more sensitive at lower gas flow rates. Adjusting the PEEP dial by half a rotation at 5 L/min can more than double the PEEP, impacting ventilation settings.

More Related Videos

Characterization of the Isolated, Ventilated, and Instrumented Mouse Lung Perfused with Pulsatile Flow
10:02

Characterization of the Isolated, Ventilated, and Instrumented Mouse Lung Perfused with Pulsatile Flow

Published on: April 29, 2011

High Speed Droplet-based Delivery System for Passive Pumping in Microfluidic Devices
10:22

High Speed Droplet-based Delivery System for Passive Pumping in Microfluidic Devices

Published on: September 2, 2009

Related Experiment Videos

Last Updated: Jun 6, 2026

Protocol for Relative Hydrodynamic Assessment of Tri-leaflet Polymer Valves
11:12

Protocol for Relative Hydrodynamic Assessment of Tri-leaflet Polymer Valves

Published on: October 17, 2013

Characterization of the Isolated, Ventilated, and Instrumented Mouse Lung Perfused with Pulsatile Flow
10:02

Characterization of the Isolated, Ventilated, and Instrumented Mouse Lung Perfused with Pulsatile Flow

Published on: April 29, 2011

High Speed Droplet-based Delivery System for Passive Pumping in Microfluidic Devices
10:22

High Speed Droplet-based Delivery System for Passive Pumping in Microfluidic Devices

Published on: September 2, 2009

Area of Science:

  • Neonatal respiratory support
  • Mechanical ventilation
  • Pediatric critical care

Background:

  • The Neopuff is a device used for neonatal resuscitation.
  • Current guidelines recommend a gas flow of 5-15 L/min for Neopuff setup.
  • The sensitivity of the positive end-expiratory pressure (PEEP) valve may vary with gas flow rates.

Purpose of the Study:

  • To investigate the sensitivity of the Neopuff's PEEP valve across different gas flow rates.
  • To determine if PEEP valve performance changes within the recommended 5-15 L/min flow range.

Main Methods:

  • Five Neopuff devices were calibrated to deliver a baseline PEEP of 5 cm H2O.
  • The number of clockwise rotations to fully occlude the PEEP valve was recorded.
  • Pressure changes were measured at quarter-rotation increments across flow rates of 5, 10, and 15 L/min.

Main Results:

  • At 5 L/min, 0.5 rotations occluded the valve, increasing pressure to 11.4 cm H2O.
  • At 10 L/min, 1.7 rotations were needed, reaching 18.4 cm H2O.
  • At 15 L/min, 3.4 rotations were required, resulting in 21.5 cm H2O.
  • Half a rotation at 5 L/min doubled PEEP (5 to 11.4 cm H2O), while at 10 L/min it increased PEEP to 7.7 cm H2O, and at 15 L/min to 6.8 cm H2O.

Conclusions:

  • The Neopuff's PEEP valve exhibits increased sensitivity at lower gas flow rates.
  • Clinicians should be aware that small adjustments to the PEEP dial at lower flows (e.g., 5 L/min) can lead to significant PEEP changes.
  • This heightened sensitivity at 5 L/min necessitates careful titration of PEEP during Neopuff use to avoid barotrauma.