Related Experiment Video
Updated: Aug 9, 2026

Measurement of Antibody Effects on Cellular Function of Isolated Cardiomyocytes
Published on: March 8, 2013
A study of the CEDIA digoxin immunoassay
1Department of Laboratory Medicine, Albert Einstein College of Medicine, Bronx, NY 10461.
The CEDIA digoxin immunoassay offers a rapid and reliable method for measuring digoxin levels. This enzyme immunoassay demonstrates excellent correlation with RIA, providing accurate results efficiently.
Area of Science:
- Clinical Chemistry
- Biotechnology
- Immunoassay Development
Background:
- Digoxin is a cardiac glycoside used to treat heart failure and arrhythmias.
- Accurate monitoring of digoxin levels is crucial due to its narrow therapeutic index.
- Existing methods like radioimmunoassay (RIA) can be time-consuming and require specialized handling.
Purpose of the Study:
- To evaluate the performance of the CEDIA digoxin immunoassay on the Cobas-Bio centrifugal analyzer.
- To assess the correlation between the CEDIA assay and RIA for digoxin quantification.
- To determine the precision, stability, and efficiency of the CEDIA assay.
Main Methods:
- The CEDIA digoxin immunoassay was performed using the Cobas-Bio centrifugal analyzer.
- Enzyme activity was measured based on the recombination of beta-galactosidase fragments.
- Assays were conducted on sera with known digoxin concentrations and compared against RIA results.
Main Results:
- Enzyme activity showed a direct proportionality to digoxin concentration.
- CEDIA results correlated well with RIA (r = 0.95), with the equation: CEDIA (microgram/L) = 1.00 x RIA - 0.06 microgram/L.
- Within-run coefficients of variation (CVs) were 1-3%, and day-to-day CVs were 3-12%.
- Reagents demonstrated stability for at least one month at 5°C.
- The assay processed samples approximately twice as fast as RIA.
Conclusions:
- The CEDIA digoxin immunoassay is a precise and accurate method for digoxin measurement.
- It offers a significant time advantage over RIA, making it suitable for high-throughput clinical laboratories.
- The assay's stability and performance on the Cobas-Bio analyzer support its clinical utility.