Related Experiment Video
Updated: Jun 6, 2026

07:22
Measuring the Motor Aspect of Cancer-Related Fatigue using a Handheld Dynamometer
Published on: February 20, 2020
New assessment of forearm strength: reliability and validity
Christopher Kevin Wong1, Neil Moskovitz
1Program in Physical Therapy, Columbia University, 710 West 168th Street, New York, NY 10032, USA. ckw7@columbia.edu
Summary
This study found that the portable Baseline hydraulic dynamometer offers good reliability for forearm strength testing. Its validity is moderate when compared to the Cybex 6000, suggesting it
Area of Science:
- Biomechanics
- Rehabilitation Engineering
- Musculoskeletal Assessment
Background:
- Accurate measurement of forearm strength is crucial for diagnosing impairments and guiding rehabilitation.
- Portable dynamometers offer convenience but require validation against established gold standards.
- The functional doorknob handle design may improve usability in clinical settings.
Purpose of the Study:
- To evaluate the reliability (intra- and interrater) of a portable hydraulic forearm dynamometer with a doorknob handle.
- To assess the criterion validity of this dynamometer by comparing its measurements to those obtained from a Cybex 6000 isometric torque assessment.
Main Methods:
- A methodological study involving 18 volunteers (30 forearms) was conducted in a single session.
- Intra- and interrater reliability were assessed for pronation and supination.
- Criterion validity was determined by comparing the hydraulic dynamometer's readings with the Cybex 6000.
Main Results:
- Excellent intrarater reliability was observed for both pronation (ICC(3,1)=.937-.961) and supination (ICC(3,1)=.923-.968).
- Good interrater reliability was found for pronation (ICC(3,2)=.927) and supination (ICC(3,2)=.847).
- Criterion validity showed moderate agreement, with correlation coefficients ranging from .574-.664 for pronation and .749-.750 for supination.
Conclusions:
- The Baseline hydraulic dynamometer with a doorknob handle demonstrates high intra- and interrater reliability.
- The device exhibits moderate validity when compared to the gold-standard Cybex 6000.
- This portable dynamometer is a reliable tool for assessing forearm strength in clinical and research settings.