Published on: February 10, 2014
Mixed-mode failure strength of implant-cement interface specimens with varying surface roughness
J Zelle1, D Janssen, S Peeters
1Orthopaedic Research Laboratory, Radboud University Nijmegen Medical Centre, P.O. Box 9101, 6500 HB Nijmegen, The Netherlands. J.Zelle@orthop.umcn.nl
Abstract:
Aseptic loosening at the implant-cement interface is a well-documented cause of failure in joint arthroplasty. Traditionally, the strength of the implant-cement interface is determined using uni-axial normal and shear loading tests. However, during functional loading, the implant fixation sites are loaded under more complex stress conditions. For this purpose, the strength of the implant-cement interface under mixed-mode tensile and shear loading conditions was determined in this study using interface specimens with varying interface roughness. For the lowest roughness value analyzed (R(a)=0.89 μm), the interface strength was 0.40-1.95 MPa at loading angles varying between pure tension and shear, whereas this was 4.90-9.90 MPa for the highest roughness value (R(a)=2.76 μm). The interface strength during pure shear (1.95-9.90 MPa) was substantially higher than during pure tension (0.58-6.67 MPa). Polynomial regression was used to fit a second-order interpolation function through the experimental interface strength data (R²=0.85; p<0.001), relating the interface strength (S [MPa]) to the interface loading angle (α [degrees]) and interface roughness (R(a) [μm]): S(α,R(a))=0.891R²(a)+0.001α²-0.189R(a)-0.064α-0.060. Finally, an interface failure criterion was derived from the interface strength measurements, describing the risk of failure at the implant-cement interface when subjected to a certain tensile and shear stress using only the interface strength in pure tensile and shear direction. The findings presented in this paper can be used in numerical models to simulate loosening at the implant-cement interface.
More Related Videos
06:36Effects of Mechanical Methods Used in Peri-implantitis Treatment on Implant Surface Decontamination and Roughness
Published on: March 14, 2025
07:42Quasistatic Mechanical Testing for Computer-Aided Design and Manufacturing Occlusal Veneers Cemented to Milled Dentin Analog Material
Published on: December 20, 2024
Related Concept Videos
Strength of Cement
For compressive strength tests, ASTM C 109-05 standards prescribe a cement-sand mix ratio of 1:2.75 and a water/cement ratio of 0.485 for making 2-inch cubes. These cubes are mixed, cast, and cured in saturated lime water at 23°C until testing. Flexural strength testing, outlined in ASTM C...
Bonding and Strength of Aggregate
Microcracking in Concrete
Porosity in Cement Paste
The balance of water to cement in the mix is critical—it...
Fineness of Cement
Direct...
Tensile Strength Considerations of Concrete
The dimensions and shape of a concrete specimen also...