Related Experiment Video
Updated: Jun 6, 2026

Measurement of X-ray Beam Coherence along Multiple Directions Using 2-D Checkerboard Phase Grating
Published on: October 11, 2016
Practical quantitative measurement of graticule misalignment relative to collimator axis of rotation
1Department of Radiation Oncology, University of California San Francisco, San Francisco, CA 94143-1708, USA. halabit@radonc.ucsf.edu
Abstract:
We design a practical procedure for measuring translational and rotational misalignment of graticule with collimator axis of rotation and collimator jaws, respectively. The procedure's quantitative results are accurate to less than 0.2 mm (at isocenter) and do not assume alignment of radiation focal spot with collimator axis of rotation. When provided with these quantitative results, the manufacturer can custom-adjust graticules to the purchaser's collimator head.