Accelerator mass spectrometry best practices for accuracy and precision in bioanalytical (14)C measurements
John S Vogel1, Jason A Giacomo, Tim Schulze-König
1Vitalea Science, 2101 Second Street, Ste. 101B, Davis, CA 95618, USA. johnsvogel@yahoo.com
Abstract:
Accelerator mass spectrometers have an energy acceleration and charge exchange between mass definition stages to destroy molecular isobars and allow single ion counting of long-lived isotopes such as (14)C (t½=5370 years.). 'Low' voltage accelerations to 200 kV allow laboratory-sized accelerator mass spectrometers instruments for bioanalytical quantitation of (14)C to 2-3% precision and accuracy in isolated biochemical fractions. After demonstrating this accuracy and precision for our new accelerator mass spectrometer, we discuss the critical aspects of maintaining quantitative accuracy from the defined biological fraction to the accelerator mass spectrometry quantitation. These aspects include sufficient sample mass for routine rapid sample preparation, isotope dilution to assure this mass, isolation of the carbon from other sample combustion gasses and use of high-efficiency biochemical separations. This review seeks to address a bioanalytical audience, who should know that high accuracy data of physiochemical processes within living human subjects are available, as long as a (14)C quantitation can be made indicative of the physiochemistry of interest.
More Related Videos
Related Concept Videos
Tandem Mass Spectrometry
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing nebulizer...
Mass Analyzers: Overview
High-Resolution Mass Spectrometry (HRMS)
Mass Analyzers: Common Types
Mass Spectrometers