In vivo assessment of coronary stents with 64-row multidetector computed tomography: analysis of metal artifacts
Giacomo D E Papini1, Filippo Casolo, Giovanni Di Leo
1Scuola di Specializzazione in Radiodiagnostica, Facoltà di Medicina e Chirurgia, Università degli Studi di Milano, Milan, Italy.
Objective:
To evaluate stent-induced artifacts by 64-row multidetector computed tomography (MDCT).
Methods:
We studied 26 stented patients with MDCT before conventional coronary angiography (CCA). The CT values were measured. Stents were classified as occluded, with significant stenosis, with nonsignificant stenosis, or patent. For the patent stents, mean in-stent and out-stent CT values were compared; stents 3 mm or smaller were compared with stents larger than 3 mm. Multidetector CT was compared with CCA.
Results:
We analyzed 42 stents. At CCA, 34 stents were patent, 5 were nonsignificantly stenosed, 1 was significantly stenosed, and 2 were occluded. At MDCT, 33 of 34 patent stents, 2 occluded stents, and 1 stent with significant stenosis were correctly diagnosed; nonsignificant stenoses were undetected, 1 patent stent was misdiagnosed as occluded (κ = 0.727). The out-stent CT value was lower than in-stent CT value both in stents 3 mm or smaller (P = 0.001) and stents larger than 3 mm (P < 0.001). The in-stent CT value of stents 3 mm or smaller was higher (P = 0.011) than that of stents larger than 3 mm.
Conclusions:
Metal artifacts cause overlooking of nonsignificant stenosis.
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...