Practical wavelength calibration considerations for UV-visible Fourier-transform spectroscopy
Applied Optics
|November 19, 2010
Summary
Laser-controlled Michelson interferometers offer excellent wavelength accuracy. Calibration with a trusted standard is crucial for correcting practical effects and achieving high precision in the UV-visible region.
Area of Science:
- Optical Physics
- Metrology
- Spectroscopy
Background:
- Modern laser-controlled Michelson interferometers possess an intrinsic wavelength scale, known as the Connes advantage, enabling high wavelength accuracy.
- Achieving a relative uncertainty of 10^-8 in line position is feasible with single-point calibration.
- Practical limitations such as finite aperture and imperfect alignment necessitate wavelength scale corrections.
Purpose of the Study:
- To discuss practical aspects of accurate interferometer calibration in the UV-visible region.
- To address critical issues in using external wavelength standards.
- To evaluate and select appropriate wavelength standards for calibration.
Main Methods:
- Utilizing a laser-controlled Michelson interferometer for wavelength calibration.
- Employing single-point, multiplicative calibration techniques.
- Investigating calibration using external wavelength standards and Fabry-Perot interferometry.
Main Results:
- Demonstrated the potential for superb wavelength accuracy (relative uncertainty ~10^-8) with proper calibration.
- Identified finite aperture and geometric misalignment as key factors requiring correction.
- Observed anomalous results for Ar II and (198)Hg I when used as wavelength standards via Fabry-Perot interferometry.
Conclusions:
- Accurate calibration is essential for realizing the full potential of Michelson interferometers for precise wavelength measurements.
- Careful selection and validation of external wavelength standards are critical for reliable calibration.
- Further investigation is needed to understand the anomalous behavior of certain potential wavelength standards.
Related Concept Videos
Ultraviolet and Visible (UV–Vis) Spectroscopy: Overview
Ultraviolet–visible (UV–visible or UV–Vis) spectroscopy is an analytical technique that investigates the interaction between matter and UV–Vis light within the electromagnetic spectrum. This method is widely used for its versatility, simplicity, and relatively quick data acquisition, making it valuable for both qualitative and quantitative analysis. When UV–Vis radiation passes through a material, molecules absorb light depending on the energy required for electronic transitions. As a result...
UV–Vis Spectrometers
The absorbance of UV and visible (UV–visible) radiations is measured using a UV–visible spectrophotometer. Deuterium lamps, which emit UV radiation, and tungsten lamps, which produce radiation in the visible region, are used as light sources in UV–visible spectrophotometers. A monochromator or prism is used for diffraction grating, i.e., to split the incoming radiation into different wavelengths. A system of slits is used to focus the desired wavelength on the sample cell. Samples for...
UV–Vis Spectroscopy: Beer–Lambert Law
The Beer-Lambert law describes the relationship between absorbance and concentration, which combines the principles established by scientists Johann Heinrich Lambert and August Beer. Lambert's law states that when light passes through a medium, the loss in intensity is directly proportional to the original intensity and the path length of the light. Beer's law proposed that the transmittance of a solution remains constant if the product of concentration and path length is constant. The modern...
Instrument Calibration
Instrument calibration is essential for ensuring that instruments produce accurate and consistent results. It is vital in manufacturing, healthcare, testing laboratories, and scientific research. Calibration processes are specific to each instrument and help enhance data accuracy. Each instrument has a unique calibration process tailored to its design and function to improve data accuracy.
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Spectrophotometry: Introduction
Spectrophotometry is the quantitative measurement of the absorption, reflection, diffraction, or transmission of electromagnetic radiation through a material as a function of the intensity and wavelength of the radiation. A spectrophotometer is a device used to measure the change in the radiation intensity caused by its interaction with the material.
The essential components of a spectrophotometer include a source of electromagnetic radiation, a slot for placing a material to be analyzed, and a...
The essential components of a spectrophotometer include a source of electromagnetic radiation, a slot for placing a material to be analyzed, and a...
UV–Vis Spectroscopy: Woodward–Fieser Rules
UV–Visible absorption spectra of conjugated dienes arise from the lowest energy π → π* transitions. The light-absorbing part of the molecule is called the chromophore, and the substituents directly attached to the chromophore are called auxochromes. A strong correlation exists between the absorption maxima, λmax, and the structure of a conjugated π system. The Woodward–Fieser rules predict the value of λmax for a given structure by adding the contributions...