Related Experiment Video
Updated: Jun 6, 2026

X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
Dose optimization in pediatric cardiac x-ray imaging
Amber J Gislason1, Andrew G Davies, Arnold R Cowen
1LXi Research, Division of Medical Physics, University of Leeds, Worsley Building, Clarendon Way, Leeds LS2 9JT, United Kingdom. a.j.gislason@leeds.ac.uk
Purpose:
The aim of this research was to explore x-ray beam parameters with intent to optimize pediatric x-ray settings in the cardiac catheterization laboratory. This study examined the effects of peak x-ray tube voltage (kVp) and of copper (Cu) x-ray beam filtration independently on the image quality to dose balance for pediatric patient sizes. The impact of antiscatter grid removal on the image quality to dose balance was also investigated.
Methods:
Image sequences of polymethyl methacrylate phantoms approximating chest sizes typical of pediatric patients were captured using a modern flat-panel receptor based x-ray imaging system. Tin was used to simulate iodine-based contrast medium used in clinical procedures. Measurements of tin detail contrast and flat field image noise provided the contrast to noise ratio. Entrance surface dose (ESD) and effective dose (E) measurements were obtained to calculate the figure of merit (FOM), CNR2/dose, which evaluated the dose efficiency of the x-ray parameters investigated. The kVp, tube current (mA), and pulse duration were set manually by overriding the system's automatic dose control mechanisms. Images were captured with 0, 0.1, 0.25, 0.4, and 0.9 mm added Cu filtration, for 50, 55, 60, 65, and 70 kVp with the antiscatter grid in place, and then with it removed.
Results:
For a given phantom thickness, as the Cu filter thickness was increased, lower kVp was favored. Examining kVp alone, lower values were generally favored, more so for thinner phantoms. Considering ESD, the 8.5 cm phantom had the highest FOM at 50 kVp using 0.4 mm of Cu filtration. The 12 cm phantom had the highest FOM at 55 kVp using 0.9 mm Cu, and the 16 cm phantom had highest FOM at 55 kVp using 0.4 mm Cu. With regard to E, the 8.5 and 12 cm phantoms had the highest FOM at 50 kVp using 0.4 mm of Cu filtration, and the 16 cm phantom had the highest FOM at 50 kVp using 0.25 mm Cu. Antiscatter grid removal improved the FOM for a given set of x-ray conditions. Under aforesaid optimal settings, the 8.5 cm phantom FOM improved by 24% and 33% for ESD and E, respectively. Corresponding improvements were 26% and 24% for the 12 cm phantom and 6% and 15% for the 16 cm phantom.
Conclusions:
For pediatric patients, using 0.25-0.9 mm Cu filtration in the x-ray beam while maintaining 50-55 kVp, depending on patient size, provided optimal x-ray image quality to dose ratios. These settings, adjusted for x-ray tube loading limits and clinically acceptable image quality, should provide a useful strategy for optimizing iodine contrast agent based cardiac x-ray imaging. Removing the antiscatter grid improved the FOM for the 8.5 and 12 cm phantoms, therefore grid removal is recommended for younger children. Improvement for the 16 cm phantom declined into the estimated margin of error for the FOM; the need for grid removal for older children would depend on practical feasibility in the clinical environment.
More Related Videos
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
11:09High-Resolution Cardiac Positron Emission Tomography/Computed Tomography for Small Animals
Published on: December 16, 2022
Related Concept Videos
Drug Dosing: Infants and Children
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Pharmacokinetics in Pediatric Patients: Drug Excretion
Imaging Studies for Cardiovascular System V: CT
X-ray Imaging
Radiological Investigation I: X-ray and CT