High Speed Sub-GHz Spectrometer for Brillouin Scattering Analysis
Published on: December 22, 2015
Humidity-induced spectral shift in a cross-dispersion echelle spectrometer and its theoretical investigation
Ke-ling Liu1, Mao Huang, Gary M Hieftje
1Institute of Process Engineering, Chinese Academy of Sciences, Beijing 100190, China. klliu@home.ipe.ac.cn
Abstract:
The relationship between ambient relative humidity H and the position shift of a spectral line was investigated both experimentally and theoretically. An echelle-based ICP emission spectrometer equipped with a CID detector was used for experimental verification of the derived model. The shift of a spectral line is quantitatively described by two defined spectral shift functions: delta lambda x(x, lamda, H) (in the x direction of the CID detector) and delta lambda y(y, lambda, H) (in the y direction of the CID detector). Experimental results indicate that delta lamda x(x, lambda, H) does not change with a variation in ambient relative humidity, but delta lamda y(y, lambda, H) does. A spectral shift equation, i.e., an empirical second-order polynomial equation, can be used to describe the relationship between delta lamda y(y, lambda, H) and H. Based on the classical dipole model, classical mechanics and electrodynamics the empirical spectral-shift equation involving delta lambda y(y, lambda, H) and H was theoretically deduced. The theoretical result is in good agreement with the experimental findings. The theoretical results indicate that the coefficients of the empirical spectral-shift equation are related to the basic physical parameters of materials and the geometric configuration of the echelle CID ICP-AES, and also provide physical meaning to the coefficients of the empirical shift equation obtained experimentally.
Related Concept Videos
NMR Spectroscopy: Chemical Shift Overview
For instance, the proton...
¹H NMR: Interpreting Distorted and Overlapping Signals
As Δν decreases and the signals move closer, the doublets appear increasingly distorted. The intensities of the inner lines increase at the cost of those of the outer lines as the signals are slanted or...
Atomic Emission Spectroscopy: Interference
IR Spectroscopy: Hooke's Law Approximation of Molecular Vibration
According to Hooke's law, the vibrational frequency is directly proportional to the...
Molecular Spectroscopy: Absorption and Emission
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used.