Clinical trials of a urethral dose measurement system in brachytherapy using scintillation detectors
Natalka Suchowerska1, Michael Jackson, Jamil Lambert
1Radiation Oncology, Royal Prince Alfred Hospital, New South Wales, Australia. natalka@email.cs.nsw.gov.au
Purpose:
To report on the clinical feasibility of a novel scintillation detector system with fiberoptic readout that measures the urethral dose during high-dose-rate brachytherapy treatment of the prostate.
Methods And Materials:
The clinical trial enrolled 24 patients receiving high-dose-rate brachytherapy treatment to the prostate. After the first 14 patients, three improvements were made to the dosimeter system design to improve clinical reliability: a dosimeter self-checking facility; a radiopaque marker to determine the position of the dosimeter, and a more robust optical extension fiber.
Results:
Improvements to the system design allowed for accurate dose measurements to be made in vivo. A maximum measured dose departure of 9% from the calculated dose was observed after dosimeter design improvements.
Conclusions:
Departures of the measured from the calculated dose, after improvements to the dosimetry system, arise primarily from small changes in patient anatomy. Therefore, we recommend that patient response be correlated with the measured in vivo dose rather than with the calculated dose.