Syringe injection pressure: a neglected factor in blood PO2 determination
Abstract:
The mean arterial blood PO2 values of 20 patients' samples taken into Sarstedt 2 mL Monovette syringes were found to be initially 0.43 kPa higher and after 15 min storage 0.93 kPa higher than the mean values using Corning 3 mL Blood Gas Kit syringes. The Sarstedt syringe allowed entry of small air bubbles and the plunger was noted to be stiff. We speculated that high injection pressures might increase measured PO2. Using each type of syringe, 20 ITU staff injected saline into a Corning 178 pH/blood analyser linked to a pressure transducer via a three-way connector. Mean injection pressures for Sarstedt and Corning syringes were 17.15 and 6.42 kPa, respectively (P less than 0.005). Following injection at pressures between 2.8 and 26.6 kPa, the mean PO2 of equilibrated saline containing an air bubble was 0.80 kPa higher than the mean value obtained at injection pressures of less than 2.8 kPa. However, with no air bubble the mean PO2 of equilibrated saline was still 0.45 kPa higher when injected at pressures between 2.8 and 26.6 kPa than the mean value obtained with injection pressures of 2.8 kPa. We suggest that for optimal PO2 determinations syringes should not only allow minimal air contamination but also have plungers that reduce injection pressure to a minimum.
More Related Videos
08:51Multiple Intravenous Bolus Dosing and Invasive Hemodynamic Assessment in a Hypoxia-Induced Mouse Pulmonary Artery Hypertension Model
Published on: November 11, 2022
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)
Pre-Procedural Guidelines for Assessing Blood Pressure
Measurement of Blood Pressure
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...