A multi-channel low-power system-on-chip for single-unit recording and narrowband wireless transmission of neural signal
Related Experiment Video
Updated: Jun 6, 2026

A Wireless, Bidirectional Interface for In Vivo Recording and Stimulation of Neural Activity in Freely Behaving Rats
Published on: November 7, 2017
A Bonfanti1, M Ceravolo, G Zambra
1Dipartimento di Elettronica e Informazione, Politecnico di Milano, piazza L. da Vinci 32, 20133, Italy. bonfanti@elet.polimi.it
Abstract:
This paper reports a multi-channel neural recording system-on-chip (SoC) with digital data compression and wireless telemetry. The circuit consists of a 16 amplifiers, an analog time division multiplexer, an 8-bit SAR AD converter, a digital signal processor (DSP) and a wireless narrowband 400-MHz binary FSK transmitter. Even though only 16 amplifiers are present in our current die version, the whole system is designed to work with 64 channels demonstrating the feasibility of a digital processing and narrowband wireless transmission of 64 neural recording channels. A digital data compression, based on the detection of action potentials and storage of correspondent waveforms, allows the use of a 1.25-Mbit/s binary FSK wireless transmission. This moderate bit-rate and a low frequency deviation, Manchester-coded modulation are crucial for exploiting a narrowband wireless link and an efficient embeddable antenna. The chip is realized in a 0.35- εm CMOS process with a power consumption of 105 εW per channel (269 εW per channel with an extended transmission range of 4 m) and an area of 3.1 × 2.7 mm(2). The transmitted signal is captured by a digital TV tuner and demodulated by a wideband phase-locked loop (PLL), and then sent to a PC via an FPGA module. The system has been tested for electrical specifications and its functionality verified in in-vivo neural recording experiments.