Related Experiment Video
Updated: Jun 6, 2026

Optimizing Minimally Invasive Spine Surgery: A Fully 3D CT O-Arm Navigated Workflow in MIS TLIF
Published on: October 17, 2025
Sensorized guidewires with MEMS tri-axial force sensor for minimally invasive surgical applications
Liang Lou1, Kotlanka Ramakrishna, Lichun Shao
1A*Star Institute of microelectronics, Science park 11, Singapore 117685.
Abstract:
This paper describes the design of a tri-axial microelectromechanical force sensor (FS) that can be mounted on the tip of the guidewire. Piezoresistive silicon nanowires (SiNW) are embedded into a cross cantilever design with a manoeuvrable stylus to allow the detection of force in all directions. The electrical resistance changes in the four SiNWs are used to decode an arbitrary force applied onto the FS. The sensitivity of the device can be improved by two orders of magnitude compared to bulk Si thanks to the giant piezoresistive effects offered by the SiNW. Robustness of the FS is improved due to the novel design by incorporating a mechanical stopper at the tip of the stylus. Finite element analysis (FEM) analysis was used in designing the FS.

