Feasibility and radiation dose of high-pitch acquisition protocols in patients undergoing dual-source cardiac CT
Wieland H Sommer1, Edda Albrecht, Fabian Bamberg
1Department of Clinical Radiology, University of Munich, Grosshadern Campus, Marchioninstr. 15, 81377, Munich, Germany. Wieland.sommer@med.uni-muenchen.de
Objective:
The objective of this study was to compare image quality and radiation dose between high-pitch and established retrospectively and prospectively gated cardiac CT protocols using an Alderson-Rando phantom and a set of patients.
Materials And Methods:
An anthropomorphic Alderson-Rando phantom equipped with thermoluminiscent detectors and a set of clinical patients underwent the following cardiac CT protocols: high-pitch acquisition (pitch 3.4), prospectively triggered acquisition, and retrospectively gated acquisition (pitch 0.2). For patients with sinus rhythm below 65 beats per minute (bpm), high-pitch protocol was used, whereas for patients in sinus rhythm between 65 and 100 bpm, prospective triggering was used. Patients with irregular heart rates or heart rates of ≥ 100 bpm, were examined using retrospectively gated acquisition. Evaluability of coronary artery segments was determined, and effective radiation dose was derived from the phantom study.
Results:
In the phantom study, the effective radiation dose as determined with thermoluminescent detector (TLD) measurements was lowest in the high-pitch acquisition (1.21, 3.12, and 11.81 mSv, for the high-pitch, the prospectively triggered, and the retrospectively gated acquisition, respectively). There was a significant difference with respect to the percentage of motion-free coronary artery segments (99%, 87%, and 92% for high-pitch, prospectively triggered, and retrospectively gated, respectively (p < 0.001), whereas image noise was lowest for the high-pitch protocol (p < 0.05).
Conclusion:
High-pitch scans have the potential to reduce radiation dose up to 61.2% and 89.8% compared with prospectively triggered and retrospectively gated scans. High-pitch protocols lead to excellent image quality when used in patients with stable heart rates below 65 bpm.
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies III: Computed Tomography