Related Experiment Video
Updated: Jun 6, 2026

Integration of Miniaturized Solid Phase Extraction and LC-MS/MS Detection of 3-Nitrotyrosine in Human Urine for Clinical Applications
Published on: July 14, 2017
Liquid chromatography-tandem mass spectrometry determination of N-nitrosamines released from rubber or elastomer
1National Institute of Food and Drug Safety Evaluation, Eungpyung-gu, Seoul, Korea.
Abstract:
A liquid chromatography-tandem mass spectrometry (LC-MS/MS) method as an alternative to a gas chromatography-thermal energy analyser (GC-TEA) method recommended by the European Committee on Standardization (CEN) was validated for the simultaneous determination of eight N-nitrosamines released into artificial saliva from rubber or elastomer teats and soothers. N-nitroso-dipropylamine-d₁₄ (NDPA-d₁₄) was used as internal standard for accurate quantification. The method was validated with relatively good analytical results, including sufficiently low limits of detection (0.1-2 µg kg⁻¹) of sample) and good linearity (r²> 0.99) throughout the studied concentration ranges. Intra- and inter-day precisions expressed with the relative standard deviation (RSD, %) were 3.4-8.0% and 4.4-11.3%, which were below the performance criteria based on one-half of the value derived from the Horwitz value. It was also found that the LC-MS/MS method is sufficiently rugged and successfully applicable to its routine analysis for the compliance test of Commission Directive 93/11/EEC.