An innovative method for determining lipemia interference in blood specimens.
Murray A Rosenthal1, Hyman B Katz
1Alfa Wassermann Diagnostic Technologies, West Caldwell, NJ, USA. mrosenthal@awdt.us
Clinica Chimica Acta; International Journal of Clinical Chemistry
|November 27, 2010
Summary
A new method accurately assesses lipemia interference in blood tests. This technique confirmed no triglyceride interference with HDL measurements up to 9.78 mmol/l on the Hitachi S40 analyzer.
Area of Science:
- Clinical Chemistry
- Biochemical Analysis
- Diagnostic Assays
Background:
- Lipemia interference in blood samples is a common issue in clinical diagnostics.
- Existing methods for detecting lipemia interference have limitations in confirming the cause.
- A novel method was evaluated to improve the assessment of lipemia interference.
Purpose of the Study:
- To assess a novel method for evaluating lipemia interference in clinical samples.
- To determine the reliability of a new technique for identifying interference caused by lipemia.
- To validate a method for assessing triglyceride interference in HDL cholesterol assays.
Main Methods:
- Matched lipemic and non-lipemic serum samples with similar HDL cholesterol were mixed.
- Samples were analyzed for HDL and triglycerides across five different concentration mixtures.
- Percent recovery of HDL was calculated for each mixture to assess interference.
Main Results:
- HDL recoveries ranged from 95.9% to 101.1% across 30 tested concentrations.
- Tested HDL concentrations were between 0.78 and 2.16 mmol/l.
- Triglycerides concentrations ranged from 1.06 to 9.78 mmol/l in the tested samples.
Conclusions:
- The novel matching method demonstrated no triglyceride interference with the HDL assay on the Hitachi S40 Clinical Analyzer up to 9.78 mmol/l.
- This simple matching method is effective for evaluating lipemia interference.
- The study confirms the accuracy of the novel method in assessing interference.