Monitoring microemboli during cardiopulmonary bypass with the EDAC quantifier
John E Lynch1, Christopher Wells, Tom Akers
1Luna Innovations Incorporated, Hampton, Virginia, USA. lyncht@lunainnovations.com
Abstract:
Gaseous emboli may be introduced into the bypass circuit both from the surgical field and during perfusionist interventions. While circuits provide good protection against massive air embolism, they do not remove gaseous microemboli (GME) from the bypass circuit. The purpose of this preliminary study is to assess the incidence of GME during bypass surgery and determine if increased GME counts were associated with specific events during bypass surgery. In 30 cases divided between 15 coronary artery bypass grafts and 15 valve repairs, GME were counted and sizedt the three locations on the bypass circuit using the EDAC" Quantifier (Luna Innovations, Roanoke, VA). A mean of 45,276 GME were detected after the arterial line filter during these 30 cases, with significantly more detected (p = .04) post filter during valve cases (mean = 72,137 +/- 22,113) than coronary artery bypass graft cases (mean = 18,416 +/- 7831). GME detected post filter were significantly correlated in time with counts detected in the venous line (p < .001). Specific events associated with high counts included the initiation of cardiopulmonary bypass, heart manipulations, insertion and removal of clamps, and the administration of drugs. Global factors associated with increased counts post filter included higher venous line counts and higher post reservoir/bubble trap counts. The mean number of microemboli detected during bypass surgery was much higher than reported in other studies of emboli incidence, most likely due to the increased sensitivity of the EDAC Quantifier compared to other detection modalities. The results furthermore suggest the need for further study of the clinical significance of these microemboli and what practices may be used to reduce GME incidence. Increased in vitro testing of the air handling capability of different circuit designs, along with more clinical studies assessing best clinical practices for reducing GME activity, is recommended.
Insights
Gaseous microemboli (GME) are common during bypass surgery, with higher incidence during valve repair than coronary artery bypass graft procedures. Specific surgical events correlate with increased GME counts, necessitating further research into their clinical significance and reduction strategies.
Area of Science:
- Cardiovascular Surgery
- Biomedical Engineering
- Perfusion Technology
Background:
- Gaseous emboli can enter bypass circuits during surgery and interventions.
- Standard bypass circuits effectively prevent massive air embolism but not gaseous microemboli (GME).
- The incidence and sources of GME during cardiopulmonary bypass remain incompletely understood.
Purpose of the Study:
- To assess the incidence of GME during bypass surgery.
- To determine if increased GME counts correlate with specific surgical events.
- To compare GME incidence between coronary artery bypass graft and valve repair procedures.
Main Methods:
- A preliminary study involving 30 bypass surgery cases (15 CABG, 15 valve repair).
- Gaseous microemboli (GME) were quantified at three circuit locations using the EDAC Quantifier.
- Correlation analysis was performed between GME counts and surgical events/factors.
Main Results:
- A mean of 45,276 GME were detected post-arterial line filter.
- Significantly higher GME counts were observed during valve repair cases compared to CABG.
- High GME counts correlated with specific events like cardiopulmonary bypass initiation, heart manipulation, and clamp application.
Conclusions:
- The study detected a higher mean GME incidence than previously reported, likely due to the EDAC Quantifier's sensitivity.
- Specific surgical events and circuit factors are associated with increased GME.
- Further research is recommended to investigate the clinical significance of GME and develop strategies for their reduction.
Related Concept Videos
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for diagnosing...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...