Cementless total hip replacement using second-generation components: a 12- to 16-year follow-up
1Mercy Medical Center, 2700 W Ninth Avenue, Suite 125, Oshkosh, Wisconsin 54904, USA. mjohnson@KChipandknee.com
The Journal of Bone and Joint Surgery. British Volume
|December 2, 2010
Summary
Second-generation uncemented total hip replacements show excellent long-term fixation. Refined component designs in these cementless hip arthroplasty procedures resulted in minimal osteolysis and loosening over 14 years.
Area of Science:
- Orthopedic Surgery
- Biomaterials Engineering
- Medical Device Design
Background:
- Cementless total hip replacement (THR) is a common orthopedic procedure.
- Second-generation implants aimed to improve fixation and longevity.
- Long-term outcomes of specific uncemented designs require continued evaluation.
Purpose of the Study:
- To evaluate the long-term clinical and radiological outcomes of second-generation uncemented total hip replacements.
- To assess the fixation, wear, and complication rates of specific titanium acetabular and femoral components.
Main Methods:
- A retrospective review of 123 uncemented total hip replacements performed between 1993-1994.
- Utilized T-Tap ST acetabular shells and Taperloc femoral components.
- Clinical and radiological follow-up was conducted for all patients at a mean of 14 years.
Main Results:
- Overall revision rate was low (4/123), with one femoral fracture and three aseptic acetabular loosening.
- No components were radiologically loose, with minimal osteolysis observed in a small number of hips.
- Mean femoral head penetration rate was 0.036 mm/year, indicating low polyethylene wear.
Conclusions:
- Refined component designs in cementless primary total hip replacement are associated with excellent long-term fixation.
- Second-generation uncemented THR components demonstrate durable performance with low complication rates.