Albuterol delivery via intrapulmonary percussive ventilator and jet nebulizer in a pediatric ventilator model
Ariel Berlinski1, J Randy Willis
1Department of Pediatrics, College of Medicine, University of Arkansas for Medical Sciences, 1 Children's Way, Little Rock, AR 72202, USA. berlinskiariel@uams.edu
Background:
In-line administration of bronchodilators is widely used in pediatric patients receiving mechanical ventilation. We compared the amount of albuterol captured at the end of the endotracheal tube (ETT) with an intrapulmonary percussive ventilator (IPV) versus a Salter 8900 jet nebulizer placed in-line in a pediatric ventilator model, under various operating conditions. We hypothesized that the type of aerosol generator, tidal volume (V(T)), and position in the ventilator circuit would influence the albuterol delivery.
Methods:
We connected a ventilator to a heated-wire ventilator circuit (heated to 37°C), to a cuffed 5.5-mm inner-diameter ETT, to a lung model, and ran the ventilator at pediatric ventilation settings: pressure-regulated volume-control mode, respiratory rate 20 breaths/min, PEEP 5 cm H(2)O, F(IO2) 0.4, inspiratory time 0.75 s, inspiratory rise time 0.15 s, and flow-triggering at 3 L/min. We collected aerosol with a filter at the distal end of the ETT (ie, between the ETT and the lung model). We tested 3 IPV units run on central O(2) at 50-PSI, with a drive pressure of 25 cm H(2)O, and 3 Salter 8900 units run on central O(2) at 6 L/min. We diluted 5 mg of albuterol in saline, and nebulized 3 mL with the jet nebulizer and 10 mL with the IPV. We studied V(T) (100 mL vs 200 mL), position in the circuit (at the humidifier vs at the Y-piece), and the IPV's "easy" and "hard" percussion settings.
Results:
When positioned at the humidifier, the IPV delivered significantly less albuterol than the jet nebulizer (3.1-3.9-fold difference, P = .002). When the IPV was moved to the Y-piece, the albuterol delivery was similar to that of the jet nebulizer at either position. Neither increasing the V(T) nor increasing the IPV settings increased the albuterol delivery.
Conclusions:
The IPV delivered less albuterol than the jet nebulizer when placed at the humidifier. IPV was equivalent to jet nebulizer when placed at the Y-piece. Doubling the V(T) did not increase aerosol delivery.
Related Concept Videos
Inhaled Medications
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation I: Indication and Settings
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...