Enabling High Grayscale Resolution Displays and Accurate Response Time Measurements on Conventional Computers
Published on: February 29, 2012
8-bit or 11-bit monochrome displays--which image is preferred by the radiologist?
Stephan Bender1, Kai Lederle, Christel Weiss
1Department of Clinical Radiology and Nuclear Medicine, University Medical Centre Mannheim, Heidelberg University, Medical Faculty Mannheim, Theodor-Kutzer-Ufer 1-3, 68167 Mannheim, Germany. mail.stephan.bender@googlemail.com
Objectives:
Vendors of medical displays promise a better diagnostic performance using 10- or 11-bit instead of 8-bit monochrome displays. We measured the gain of "Just Noticeable Difference" (JND) steps using high grayscale resolutions and evaluated the preference of radiologists regarding different aspects of image quality.
Methods:
The amount of JND steps was measured on a display using 8-, 10- and 11-bit Look Up Tables (LUT). Radiological images were presented simultaneously using an 8-bit and an 11-bit LUT, radiologists where asked to determine the better image presentation regarding quality parameters like sharpness, contrast and detectability of details.
Results:
The 10-bit technology realized more than twice the number of JND steps compared to an 8-bit LUT, an 11-bit LUT presented only few additional JND steps. The radiologists did not attest the higher grayscale resolution a better image quality, they regarded the 8-bit technology to show a better sharpness and contrast, although this had no impact on the detectability of details.
Conclusions:
Our measurements confirmed that the higher grayscale resolution results in a more complete visualization of image information. But radiologists partially judged this as a lack of sharpness and contrast and generally preferred the 8-bit display.
Related Concept Videos
X-ray Imaging
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Radiological Investigation I: X-ray and CT
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...