Estimates of total analytical error in consumer and hospital glucose meters contributed by hematocrit, maltose, and
Martha E Lyon1, Jeffrey A DuBois, Gordon H Fick
1Department of Pathology and Laboratory Medicine, University of Calgary, and Alberta Children's Hospital, Calgary, Alberta, Canada. martha.lyon@cls.ab.ca
Background:
Patients and physicians expect accurate whole blood glucose monitoring even when patients are anemic, are undergoing peritoneal dialysis, or have slightly elevated ascorbate levels. The objective of this study was to estimate analytical error in two consumer and two hospital glucose meters contributed by variations in hematocrit, maltose, ascorbate, and imprecision.
Methods:
The influence of hematocrit (20-60%), maltose, and ascorbate were tested alone and in combination with each glucose meter and with a reference plasma glucose method at three concentrations of glucose. Precision was determined by consecutive analysis (n=20) at three levels of glucose. Multivariate regression analysis was used to estimate the bias associated with the interferences, alone and in combination. Total analytical error was estimated as |% bias|+1.96 (% imprecision).
Results:
Three meters demonstrated hematocrit bias that was dependent upon glucose concentration. Maltose had profound concentration-dependent positive bias on the consumer meters, and the extent of maltose bias was dependent on hematocrit. Ascorbate produced small but statistically significant biases on three meters. Coincident low hematocrit, presence of maltose, and presence of ascorbate increased the observed bias and was summarized by estimation of total analytical error. Among the four glucose meter devices assessed, estimates of total analytical error in glucose measurement ranged from 6 to 68% under the conditions tested.
Conclusions:
The susceptibility of glucose meters to clinically significant analytical biases is highly device-dependent, and low hematocrit exacerbated the observed analytical error.
Related Concept Videos
Systematic Error: Methodological and Sampling Errors
Sampling errors originate from improper sampling methods or the wrong sample population. These errors can be minimized by refining the sampling strategy. Defective instruments or faulty calibrations are the sources of instrumental...
Errors occurring during blood pressure monitoring
Several factors...
Uncertainty in Measurement: Reading Instruments
Accuracy and Errors in Hypothesis Testing
In hypothesis testing, the probability of making a Type I error, denoted as α, is commonly set at 0.05. This significance level indicates a 5% chance...
Hyperglycemia
Contaminants and Errors
Another key consideration is determining the appropriate number of samples required to...