Pharyngeal flow interval: a novel impedance-based parameter correlating with aspiration

L Noll1, N Rommel, G P Davidson

  • 1Gastroenterology Unit, Child, Youth & Women's Health Service, North Adelaide, SA, Australia.

Insights

Pharyngeal impedance recording, using a novel flow interval parameter, can identify aspiration risk in children with cerebral palsy. This method shows promise for predicting swallowing difficulties.

Area of Science:

  • Otolaryngology
  • Swallowing Disorders
  • Biomedical Engineering

Background:

  • Pharyngeal impedance recording is an emerging technique for assessing swallowing function.
  • Its utility in evaluating pharyngeal function, particularly in relation to bolus residue and aspiration, requires further establishment.
  • This study investigates a novel parameter, impedance flow interval, and the impact of catheter configuration.

Purpose of the Study:

  • To evaluate the impedance flow interval as a novel parameter for assessing pharyngeal function.
  • To correlate impedance flow interval with post-swallow bolus residue and deglutitive aspiration.
  • To assess the effect of different catheter configurations on impedance measurements.

Main Methods:

  • 12 children (1.8-13.5 years) with cerebral palsy underwent simultaneous videofluoroscopy and pharyngeal impedance recording.
  • Two catheter configurations (1.9mm diameter/1cm electrodes and 3.2mm diameter/2cm electrodes) were utilized.
  • Impedance flow interval was calculated and correlated with fluoroscopic findings of residue and aspiration.

Main Results:

  • A longer flow interval was significantly associated with aspiration across both catheter configurations.
  • A longer flow interval was associated with bolus residue, particularly with the larger catheter (Catheter 2).
  • Higher flow interval was a stronger predictor of aspiration (OR 13.4) than the presence of residue (OR 3.8).

Conclusions:

  • Impedance measurement, specifically the flow interval, can detect alterations in pharyngeal swallow flow characteristics.
  • This novel parameter shows potential for predicting deglutitive aspiration risk in pediatric patients.
  • Pharyngeal impedance recording offers a promising, objective method for swallowing function assessment.
Abstract

Related Concept Videos

Suctioning the Oropharyngeal Airway01:25

Suctioning the Oropharyngeal Airway

In preparing for oropharyngeal airway suctioning, a nurse must gather all necessary equipment, including a suction unit with tubing, a prepackaged suction kit, sterile gloves, water or saline for irrigation, a water-soluble lubricant, and additional personal protective equipment (such as a gown, mask, and goggles) to control infections.
After assembling the equipment, the nurse should practice hand hygiene and don appropriate PPE according to infection control guidelines to avoid the...
Suctioning the Nasopharyngeal Airway01:29

Suctioning the Nasopharyngeal Airway

Nasopharyngeal suctioning is a procedure to remove secretions from the upper part of the respiratory tract that the patient cannot clear independently. It helps maintain airway patency and prevents complications such as aspiration pneumonia.
Equipment Required
Application of Integration: Problem Solving01:30

Application of Integration: Problem Solving

The process of breathing involves the periodic intake and expulsion of air, known as the respiratory cycle, which typically lasts about five seconds. Modeling the volume of air inhaled into the lungs as a function of time provides insight into both the dynamics and efficiency of pulmonary ventilation. This volume is determined by integrating the airflow rate over time, which captures the cumulative effect of air entering the lungs.Sinusoidal Model of AirflowAirflow during respiration is not...
Assessment of Ventilation II: Respiratory Depth and Rhythm01:29

Assessment of Ventilation II: Respiratory Depth and Rhythm

Respiratory Depth
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Urodynamic Studies: Uroflowmetry01:19

Urodynamic Studies: Uroflowmetry

Uroflowmetry is a non-invasive urodynamic test designed to measure various aspects of urination, including volume, flow rate, and the time to void. This test is crucial for diagnosing and assessing conditions such as bladder outlet obstruction, bladder dysfunction, incomplete bladder emptying, incontinence, and urinary tract blockages caused by benign prostatic hyperplasia (BPH) and urethral strictures.Pre-Test Instructions:Before a uroflowmetry test, patients are typically advised to drink...
Anatomy of Respiratory System I: Upper Respiratory Tract01:29

Anatomy of Respiratory System I: Upper Respiratory Tract

The upper respiratory tract plays a vital role in the respiratory system, comprising several structures that facilitate air intake and prepare air for the lungs. It also serves as the first line of defense against pathogens and particles. This tract includes the nose and nasal cavity, the oral cavity, the paranasal sinuses, and the pharynx, each with specific functions and features.
Nose and nasal cavity
The nose and nasal cavity represent the main external openings of the respiratory tract.