Related Experiment Video
Updated: Jun 6, 2026

Minimally Invasive Treatment for Thoracolumbar Burst Fracture Using Sagittal Alignment Screws and A Trauma Reduction Device
Published on: November 8, 2024
Lateral mass screw complications: analysis of 1662 screws
Pavlos Katonis1, Stamatios A Papadakis, Spyros Galanakos
1Department of Orthopaedics, Crete University, Medical School, Herakleion, Athens, Greece.
Study Design:
Retrospective, consecutive patient series.
Objective:
To quantify the risks and the complications associated with screw fixation devices of the cervical spine.
Summary Of Background Data:
The usefulness of lateral mass internal fixation has been well documented in the clinical setting. However, there is a paucity of studies examining the complications associated with these devices in a degenerative clinical setting.
Methods:
From 1999 to 2007, 225 consecutive patients underwent posterior cervical fixation using a screw-plate and polyaxial screw-rod implant systems. There were 105 women and 120 men (age range: 45 to 84 y; mean, 68 y). In all patients, the surgical indication was cervical spondylosis with myelopathy. Mean follow-up interval was 18 months (range: 12 to 72 mo). Screw position was evaluated by computed tomography scanning postoperatively in all patients. Clinical and radiographic outcome was assessed at each visit after surgery.
Results:
Intraoperative complications include fracture of lateral mass in 27 screws placement and nerve irritation in 3 bicortical screws. Early complications include hematoma formation in 2 cases and C5 root palsy in 5 cases after spinal canal decompression. Late complications include pseudarthrosis in 6 cases and screw pull-out in 3 cases. There were no cases of spinal cord or vertebral artery injury, infections, deaths, or adjacent segment disease. All patients had radiographic union, and no patient developed mechanical implant failure requiring removal of instrumentation. Reoperation was required in 14 (6.2%) cases because of nerve injury, hematoma formation, pseudarthrosis, and screw pull-out.
Conclusions:
Our clinical findings indicate that lateral mass fixation can be used safely with minimal complications and low rate of morbidity for cervical myelopathy treatment.
Related Concept Videos
Frictional Forces on Screws
A jack with a square-threaded screw is a mechanical device used to lift heavy loads by applying a force at its handle. When the force is applied, the screw turns, raising the load. The screw can...
Self-Locking Screw
A square-threaded screw jack carrying a load is considered self-locking if the screw retains its position even after the moment applied to it is removed.
Screw: Problem Solving
To evaluate the...
General Case of Eccentric Axial Loading
Consider a member subjected to equal and opposite forces that are applied along a line that does not coincide with the member's neutral axis. In unsymmetrical bending,...