Continuous Venous-Arterial Doppler Ultrasound During a Preload Challenge
Published on: January 20, 2023
Arterial pressure-based cardiac output monitoring: a multicenter validation of the third-generation software in
Daniel De Backer1, Gernot Marx, Andrew Tan
1Department of Intensive Care, Erasme University Hospital, Université Libre de Bruxelles, Route de Lennik 808, 1070 Brussels, Belgium. ddebacke@ulb.ac.be
Purpose:
Second-generation FloTrac software has been shown to reliably measure cardiac output (CO) in cardiac surgical patients. However, concerns have been raised regarding its accuracy in vasoplegic states. The aim of the present multicenter study was to investigate the accuracy of the third-generation software in patients with sepsis, particularly when total systemic vascular resistance (TSVR) is low.
Methods:
Fifty-eight septic patients were included in this prospective observational study in four university-affiliated ICUs. Reference CO was measured by bolus pulmonary thermodilution (iCO) using 3-5 cold saline boluses. Simultaneously, CO was computed from the arterial pressure curve recorded on a computer using the second-generation (CO(G2)) and third-generation (CO(G3)) FloTrac software. CO was also measured by semi-continuous pulmonary thermodilution (CCO).
Results:
A total of 401 simultaneous measurements of iCO, CO(G2), CO(G3), and CCO were recorded. The mean (95%CI) biases between CO(G2) and iCO, CO(G3) and iCO, and CCO and iCO were -10 (-15 to -5)% [-0.8 (-1.1 to -0.4) L/min], 0 (-4 to 4)% [0 (-0.3 to 0.3) L/min], and 9 (6-13)% [0.7 (0.5-1.0) L/min], respectively. The percentage errors were 29 (20-37)% for CO(G2), 30 (24-37)% for CO(G3), and 28 (22-34)% for CCO. The difference between iCO and CO(G2) was significantly correlated with TSVR (r(2) = 0.37, p < 0.0001). A very weak (r(2) = 0.05) relationship was also observed for the difference between iCO and CO(G3).
Conclusions:
In patients with sepsis, the third-generation FloTrac software is more accurate, as precise, and less influenced by TSVR than the second-generation software.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Assessment of blood pressure in brachial artery(two-step method)
Measurement of Blood Pressure