The change of rotational freedom following different insertion torques in three implant systems with implant driver
Joo-Hyun Kwon1, Chong-Hyun Han, Sun-Jai Kim
1Clinical Assistant Professor, Department of Prosthodontics, YongDong Severance Dental Hospital, College of Dentistry, Yonsei University, Seoul, Korea.
Statement Of Problem:
Implant drivers are getting popular in clinical dentistry. Unlike to implant systems with external hex connection, implant drivers directly engage the implant/abutment interface. The deformation of the implant/abutment interface can be introduced while placing an implant with its implant driver in clinical situations.
Purpose:
This study evaluated the change of rotational freedom between an implant and its abutment after application of different insertion torques.
Material And Methods:
Three kinds of internal connection implants were utilized for the current study (4.5 × 12 mm Xive, 4.3 × 11.5 mm Inplant Magicgrip, 4.3 × 12 mm Implantium MF). An EstheticBase, a 2-piece top, a Dual abutment was used for its corresponding implant system. The rotational freedom between an implant and its abutment were measured before and after applying 45, 100 Ncm insertion torque. Repeated measures ANOVA was used for statistical analysis.
Results:
Under 45 Ncm insertion torque, the rotational freedom between an implant and its abutment was significantly increased in Xive (P = .003). However, no significant change was noted in Inplant Magicgrip and Implantium MF. Under 100 Ncm torque, both in Xive (P = .0005) and Implatium MF (P = .03) resulted in significantly increased rotational freedom between the implant and its abutment.
Discussion:
The design of the implant/implant driver interface effectively prevented the deformation of implant/abutment interface. Little change was noted in the rotational freedom between an implant and its abutment, even though the insertion torque was far beyond clinical application.
Conclusions:
The implant/abutment joint of internally connecting implants were quite stable under insertion torque in clinical situation.
Related Concept Videos
Angle of Twist: Problem Solving
Residual Stresses in Circular Shafts
Angle of Twist - Elastic Range
Support Reactions in Three Dimensions
Ball and Socket Joint is one of the supports allowing free rotation about any axis. This freedom of rotation is...
Screw: Problem Solving
To evaluate the...
Deformation in a Circular Shaft