GC determination of N-nitrosamines by supersonic molecular beam MS equipped with triple quadrupole analyzer,
Voloshenko Anna1, Shelkov Rimma, Ovadia Lev
1The Casali Institute of Applied Chemistry, The Institute of Chemistry, The Hebrew University of Jerusalem, Jerusalem 91904, Israel.
Abstract:
The determination of 14 N-nitrosamines by a supersonic molecular beam electron ionization mass spectrometer equipped with triple quadruple analyzer, GC/SMB/EI/QQQ/MS is presented. The supersonic molecular beam electron ionization ion source allows the elucidation of the molecular ion of 13 out of the 14 examined nitrosamines (except for diphenylnitrosamine which was degraded before the analysis). It was possible to use the molecular ions of all the nitrosamines as the parent ions for multiple reactions monitoring mode, which in turn allows significant increase of specificity and lowering of the method limit of detection of the higher molecular weight nitrosamines. The instrumental LOD for different N-nitrosamines was 1-5 pg injection(-1). The proposed method was exemplified by analysis of N-nitrosamines and N-nitrosatables in rubber teats according to the British Standard BS EN 12868:1999.
Related Concept Videos
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
2° Amines to N-Nitrosamines: Reaction with NaNO2
Tandem Mass Spectrometry
Mass Spectrometry of Amines
Mass Analyzers: Common Types