High-flow nasal cannula: impact on oxygenation and ventilation in an acute lung injury model
Meg Frizzola1, Thomas L Miller, Maria Elena Rodriguez
1Nemours Children's Clinic of Wilmington, The Nemours Foundation, Alfred I. duPont Children's Hospital, Primary Research Institution, Wilmington, Delaware 19803, USA. mfrizzol@nemours.org
Introduction:
High-flow nasal cannula therapy (HFNC) has been shown to be more effective than continuous positive airway pressure (CPAP) in reducing intubations and ventilator days. HFNC likely provides mechanisms to support respiratory efficiency beyond application of distending pressure. We reason that HFNC washout of nasopharyngeal dead space impacts CO(2) removal along with oxygenation. The aim of this study was to demonstrate the flow dependence of CO(2) reduction and improved oxygenation during HFNC and the dependence on leak around the nasal prongs.
Materials And Methods:
Neonatal piglets (n=13; 2-6 kg) were injured with IV oleic acid and supported with HFNC at 2 through 8 L/min. High and low leak around the nasal prongs was accomplished by using single and double prong cannulae, respectively. Measurement of hemodynamic, respiratory and blood gas parameters were made at each setting following 10 min for physiologic equilibration. Tracheal pressures were recorded by transmural catheters.
Results:
With HFNC, CO(2) trended downward in a flow-dependent manner independent of leak. Oxygenation and tracheal pressures increased in a flow-dependent manner with the greatest effect during double prong. At 8 L/min, tracheal pressures did not exceed 6 ± 1 cmH(2) O.
Conclusions:
HFNC improves gas exchange in a flow-dependent manner; double prong had greater impact on O(2;) single prong had greater impact on CO(2) elimination.
More Related Videos
09:36Halogenated Agent Delivery in Porcine Model of Acute Respiratory Distress Syndrome via an Intensive Care Unit Type Device
Published on: September 24, 2020
06:22Surfactant Depletion Combined with Injurious Ventilation Results in a Reproducible Model of the Acute Respiratory Distress Syndrome (ARDS)
Published on: April 7, 2021
Related Concept Videos
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils, connected...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Acute Respiratory Failure-II
The underlying physiological abnormalities that contribute to hypoxemic respiratory failure include:
Acute Respiratory Failure-V
Ensure that patients are monitored continuously for their response to therapy, including changes in...
Acute Respiratory Failure-I
Definition: It is defined by specific criteria based on blood gas measurements. Hypoxemia happens when the partial pressure of oxygen (PaO2) falls below 60 mmHg. At the same time,...