Related Experiment Video
Updated: Jun 5, 2026

Mouse Body Temperature Measurement Using Infrared Thermometer During Passive Systemic Anaphylaxis and Food Allergy Evaluation
Published on: September 14, 2018
Limitations of infrared ear temperature measurement in clinical practice
Raphael Twerenbold1, Andreas Zehnder, Tobias Breidthardt
1Department of Internal Medicine, University Hospital Basel, Petersgraben 4, 4031 Basel, Switzerland. rtwerenbold@uhbs.ch
Principles:
Detection of elevated body temperature is critical in the early diagnosis of sepsis. Due to its convenience, infrared ear temperature measurement (IETM) has become the standard of care. Unfortunately, the limitations of this method are largely unexplored.
Objective:
To evaluate potential limitations of IETM, including the presence of cerumen on otoscopy, depth of penetration, side of measurement, and the impact of acclimatisation to room temperature.
Methods:
In this prospective cohort study, 333 patients presenting to the medical emergency department underwent serial IETM before and after otoscopy and cleaning of the external auditory canal. The primary endpoint was defined as mean change in infrared ear temperature (IET) before and after removal of cerumen. We also tested for the effect of penetration depth, side of measurement and impact of acclimatisation.
Results:
Otoscopy revealed cerumen in 98 patients (29%). Cerumen had a weak but statistically significant impact on IETM. The removal of cerumen obturans resulted in a rise in IET of 0.20 °C (95% CI 0.10-0.28 °C, P = 0.03). The effects of penetration depth (P = 0.39), side of measurement (P = 0.78) and impact of acclimatisation (P = 0.82) were not significant.
Conclusions:
Cerumen has a statistically significant, albeit not clinically meaningful, influence on IETM. Thus routine ear inspection prior to the use of IETM is not warranted. IETM provides highly reproducible assessments of IET irrespective of penetration depth, side of measurement and acclimatisation.
Related Concept Videos
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's forehead...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...