Related Experiment Video
Updated: Jun 5, 2026

High-throughput Detection Method for Influenza Virus
Published on: February 4, 2012
Detection of two different influenza A viruses using a nitrocellulose membrane and a magnetic biosensor
Hyo-Bong Hong1, Hans-Joachim Krause, Ki-Bong Song
1Electronics and Telecommunications Research Institute, Daejeon, Republic of Korea. hb8868@etri.re.kr
Abstract:
Here we describe a new analytical method for the detection of two influenza A viruses by nitrocellulose membrane and magnetic sensors that employ a special frequency mixing technique. The combination of the nitrocellulose membrane and magnetic bead detection permits a rapid assay procedure and excludes two steps (the development of color and the stop reaction) required for usual immunochemical detection methods such as ELISA. Quantitative virus detection was performed using magnetic beads conjugated with secondary antibody. The results were compared with conventional assay methods and with a dot-blot assay with fluorescence compound (FITC). Under optimum conditions, our new assay procedure is capable of detecting picograms of virus per well. This new method combining the nitrocellulose membrane and magnetic bead detection reduces analytical time and allows stable and repeatable analyses of samples in point-of-care applications.