Background:
Blood pressure (BP) is one of the major vital parameters monitored in the stroke unit. The accuracy of indirect BP measurement is strongly influenced by the position of both patient and arm during the measurement. Acute stroke patients are often nursed in lateral decubitus positions. The effect of these alternating body positions in relation to affected body side on the outcome and reliability of BP readings in acute stroke patients is unknown.
Methods:
An automatic oscillometric BP device was used. BP was measured in both arms in the (back) supine and both lateral decubitus positions.
Results:
In total, 54 consecutive acute stroke patients were included. Thirty-five patients had right-sided deficits and 19 patients had left-sided deficits. Supine BP readings were similar in the right and left arms regardless of side of deficit. Measurements of BP in the lateral decubitus positions resulted in significantly lower BP readings in the uppermost arm (around 12 mm Hg in both arms) and significantly higher readings in the right lowermost arm (around 6 mm Hg) compared to the supine position. This effect seemed less pronounced when the left lowermost arm was measured. There was no relation between change of BP readings in various lateral positions and side of stroke.
Conclusions:
Alternating lateral decubitus positions according to nursing standards in acute stroke patients lead to a mean 18 mm Hg BP fluctuation. This may largely be explained by hydrostatic pressure effects, partly by anatomic factors in the left lowermost arm, but not by the side of stroke.
More Related Videos
Related Concept Videos
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Alterations in Blood Pressure
Hypertension (High blood pressure)
Hypertension occurs when blood pressure readings consistently exceed the normal range. It is diagnosed when systolic blood pressure (the top number, indicating pressure while the heart beats)...
Pre-Procedural Guidelines for Assessing Blood Pressure
Blood Pressure Imbalances and Circulatory Shock
Blood Pressure: Hypertension and Hypotension
Normal blood pressure is 120/80 mm Hg. Elevated blood pressure is 120-129/under 80 mm Hg. Hypertension, warranting treatment at 130/80 mm Hg, is often asymptomatic and can lead to severe cardiovascular events, aneurysms, peripheral arterial disease, chronic renal disease, or cardiac...
Errors occurring during blood pressure monitoring
Several factors...
Measurement of Blood Pressure