Objectives:
To adapt an animal model of acute lung injury for use as a standard protocol for a screening initial evaluation of limited function, or "surge," ventilators for use in mass casualty scenarios.
Design:
Prospective, experimental animal study.
Setting:
University research laboratory.
Subjects:
Twelve adult pigs.
Interventions:
Twelve spontaneously breathing pigs (six in each group) were subjected to acute lung injury/acute respiratory distress syndrome via pulmonary artery infusion of oleic acid. After development of respiratory failure, animals were mechanically ventilated with a limited-function ventilator (simplified automatic ventilator [SAVe] I or II; Automedx, Germantown, MD) for 1 hr or until the ventilator could not support the animal. The limited-function ventilator was then exchanged for a full-function ventilator (Servo 900C; Siemens-Elema, Solna, Sweden).
Measurements And Main Results:
Reliable and reproducible levels of acute lung injury/acute respiratory distress syndrome were induced. The SAVe I was unable to adequately oxygenate five animals with Pao2 (52.0±11.1 torr) compared to the Servo (106.0±25.6 torr; p=.002). The SAVe II was able to oxygenate and ventilate all six animals for 1 hr with no difference in Pao2 (141.8±169.3 torr) compared to the Servo (158.3±167.7 torr).
Conclusions:
We describe a novel in vivo model of acute lung injury/acute respiratory distress syndrome that can be used to initially screen limited-function ventilators considered for mass respiratory failure stockpiles and that is intended to be combined with additional studies to definitively assess appropriateness for mass respiratory failure. Specifically, during this study we demonstrate that the SAVe I ventilator is unable to provide sufficient gas exchange, whereas the SAVe II, with several more functions, was able to support the same level of hypoxemic respiratory failure secondary to acute lung injury/acute respiratory distress syndrome for 1 hr.
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...