Techniques for the Analysis of Extracellular Vesicles Using Flow Cytometry
Published on: March 17, 2015
Detection of circulating microparticles by flow cytometry: influence of centrifugation, filtration of buffer, and
Emily Dey-Hazra1, Barbara Hertel, Torsten Kirsch
1Division of Nephrology, Department of Medicine, Hannover Medical School, Hannover, Germany. emily.deyhazra@googlemail.com
Abstract:
The clinical importance of microparticles resulting from vesiculation of platelets and other blood cells is increasingly recognized, although no standardized method exists for their measurement. Only a few studies have examined the analytical and preanalytical steps and variables affecting microparticle detection. We focused our analysis on microparticle detection by flow cytometry. The goal of our study was to analyze the effects of different centrifugation protocols looking at different durations of high and low centrifugation speeds. We also analyzed the effect of filtration of buffer and long-term freezing on microparticle quantification, as well as the role of Annexin V in the detection of microparticles. Absolute and platelet-derived microparticles were 10- to 15-fold higher using initial lower centrifugation speeds at 1500 × g compared with protocols using centrifugation speeds at 5000 × g (P < 0.01). A clear separation between true events and background noise was only achieved using higher centrifugation speeds. Filtration of buffer with a 0.2 μm filter reduced a significant amount of background noise. Storing samples for microparticle detection at -80°C decreased microparticle levels at days 28, 42, and 56 (P < 0.05 for all comparisons with fresh samples). We believe that staining with Annexin V is necessary to distinguish true events from cell debris or precipitates. Buffers should be filtered and fresh samples should be analyzed, or storage periods will have to be standardized. Higher centrifugation speeds should be used to minimize contamination by smaller size platelets.
Insights
Standardizing microparticle measurement is crucial. Higher centrifugation speeds and Annexin V staining improve accuracy, while buffer filtration and fresh sample analysis are recommended for reliable flow cytometry detection.
Area of Science:
- Hematology
- Analytical Chemistry
- Biotechnology
Background:
- Microparticles (MPs) from blood cells are clinically significant but lack standardized measurement methods.
- Few studies address analytical and preanalytical variables impacting MP detection.
- Flow cytometry is a key technique for MP analysis.
Purpose of the Study:
- To analyze the impact of different centrifugation protocols on MP detection.
- To evaluate the effects of buffer filtration, long-term freezing, and Annexin V staining on MP quantification.
Main Methods:
- Flow cytometry was used for MP detection.
- Variable centrifugation speeds (1500 × g vs. 5000 × g) were tested.
- Effects of 0.2 μm buffer filtration, -80°C storage, and Annexin V staining were assessed.
Main Results:
- Lower centrifugation speeds (1500 × g) yielded 10- to 15-fold higher MP counts compared to higher speeds (5000 × g).
- Higher centrifugation speeds improved separation of true events from background noise.
- Buffer filtration reduced background noise; -80°C storage decreased MP levels over time; Annexin V is essential for distinguishing true MPs from debris.
Conclusions:
- Standardized protocols using higher centrifugation speeds are necessary to minimize contamination and improve MP detection accuracy.
- Buffer filtration and Annexin V staining are recommended for reliable MP quantification.
- Fresh samples are preferred, or storage conditions must be strictly standardized for consistent results.