Laryngeal Mask Airway (LMA) Placement in a Neonatal Patient Simulator Using a Non-Inflatable Supraglottic Airway (SGA)
Published on: July 14, 2023
Performance of supraglottic airway devices and 12 month skill retention: a randomized controlled study with manikins
Henrik Fischer1, Eva Hochbrugger, Andrea Fast
1Department of Anaesthesia, General Intensive Care and Pain Control, Division of Cardiothoracic and Vascular Anaesthesia and Intensive Care, Medical University Vienna, Waehringer Guertel 18-20, 1090 Vienna, Austria. henrik.fischer@meduniwien.ac.at
Purpose:
Airway management for successful ventilation by laypersons and inexperienced healthcare providers is difficult to achieve. Bag-valve mask (BVM) ventilation requires extensive training and is performed poorly. Supraglottic airway devices (SADs) have been successfully introduced to clinical resuscitation practice as an alternative. We evaluated recently introduced (i-gel™ and LMA-Supreme™) and established SADs (LMA-Unique™, LMA-ProSeal™) and BVM used by laypeople in training sessions on manikins.
Methods:
In this randomized controlled study, 267 third-year medical students participated with informed consent and IRB approval. After brief standardized training, each participant applied all devices in a randomized order. Success of device application and ventilation was recorded. Without further training, skill retention was assessed in the same manner 12 months later. Outcome parameters were the number of application attempts, application time, tidal volume and gastric inflation rate recorded at successful attempts, and subjective ease-of-use rating by the participants.
Results:
i-gel™ and LMA-Supreme™ were the most successful in the first attempt at both assessments and in the subjective ease-of-use rating. The shortest application time was found with BVM (8 ± 5s in 2008 vs. 9 ± 5s in 2009) and i-gel (10 ± 3s vs. 12 ± 5s). Tidal volumes were disappointing with no device reaching 50% volume within the recommended range (0.4-0.6L). Gastric inflation rate was highest with BVM (18% vs. 20%) but significantly lower with all SADs (0.4-6%; p < 0.001 for 2008 and 2009).
Conclusion:
SADs showed clear advantages over BVM. Compared with LMA-Unique™ and LMA-ProSeal™, i-gel™ and LMA-Supreme™ led to higher first-attempt success rates and a shorter application time.
Related Concept Videos
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Cardiopulmonary Resuscitation II: ACLS Airway Management
Suctioning the Oropharyngeal Airway
After assembling the equipment, the nurse should practice hand hygiene and don appropriate PPE according to infection control guidelines to avoid the...
Tracheostomy Decannulation
Description of the Procedure
Decannulation refers to the permanent removal of the tracheostomy tube, signaling the resolution of the condition that initially necessitated the tracheostomy. The process requires a well-coordinated interplay between...
Suctioning the Nasopharyngeal Airway
Equipment Required
Tracheostomy Care II: Procedure
Step 1: Perform hand hygiene, and put on personal protective equipment: gown, gloves, mask and...