Related Experiment Video
Updated: Jun 5, 2026

Software-Assisted Quantitative Measurement of Osteoarthritic Subchondral Bone Thickness
Published on: March 18, 2022
Results of the EURADOS extremity dosemeter intercomparison 2009
H Stadtmann1, T W M Grimbergen, M Figel
1Seibersdorf Labor GmbH, Radiation Safety and Applications, Dosimetry, Seibersdorf, Austria. hannes.stadtmann@seibersdorf-laboratories.at
Abstract:
This paper presents the results of an intercomparison for extremity dosemeters organised by the European radiation dosimetry group in 2009. In total, 59 systems were tested during this exercise including ring, stall and wrist dosemeters. A total of 1652 dosemeters were irradiated in the selected fields of photons and beta radiation qualities on appropriate phantoms (ISO finger and pillar phantom) in the dose quantity H(p)(0.07). All irradiations were carried out in selected accredited reference dosemetry laboratories (Seibersdorf Laboratories, Austria and IRSN, France). The results show that, especially at low-energy beta radiations ((85)Kr) and for beta irradiations with large angles of incidence (60°), many tested systems show pronounced under responses. On the other hand, for photon irradiations down to energies of 16 keV most systems showed good results. A participants meeting was held at IM2010 with discussion on both general aspects of this intercomparison and specific problems.
