Evaluating the performance of the Harshaw™ DXTRAD LiF:Mg,Cu,P finger ring with two different thickness filters
S T Baker1, G H Roberts, P J Gilvin
1Health Protection Agency, Chilton, Didcot, Oxon OX11 0RQ, UK. sean.baker@hpa.org.uk
Abstract:
Amongst the dosemeters offered by the Health Protection Agency (HPA) Personal Dosimetry Service are finger rings, which use the Harshaw™ DXTRAD element. This paper describes restricted-type testing carried out by HPA on the previously untried combination of the LiF:Mg,Cu,P material and a thicker filter (42 mg cm(-2)). The tests were based on ISO 12794 [International Organization for Standardization. ISO 12794:2000 nuclear energy--radiation protection--individual thermoluminescence dosemeters for the extremities and eyes. ISO (2000)] and included energy and angle dependence of response for photons. The conclusion was, for photon dosimetry applications the thicker filter is acceptable for use in the HPA service.