Multicenter analytical evaluation of a high-sensitivity troponin T assay
A K Saenger1, R Beyrau, S Braun
1Department of Laboratory Medicine and Pathology, Hilton 3, Mayo Clinic, 200 First St. SW, Rochester, MN 55905, USA. saenger.amy@mayo.edu
Background:
High-sensitivity cardiac troponin assays are being introduced clinically for earlier diagnosis of acute myocardial infarction (AMI). We evaluated the analytical performance of a high-sensitivity cardiac troponin T assay (hscTnT, Roche Diagnostics) in a multicenter, international trial.
Methods:
Three US and 5 European sites evaluated hscTnT on the Modular® Analytics E170, cobas® 6000, Elecsys 2010, and cobas® e 411. Precision, accuracy, reportable range, an inter-laboratory comparison trial, and the 99th percentile of a reference population were assessed.
Results:
Total imprecision (CVs) were 4.6-36.8% between 3.4 and 10.3 ng/L hscTnT. Assay linearity was up to 10,000 ng/L and the limit of blank and detection were 3 and 5 ng/L, respectively. The 99th percentile reference limit was 14.2 ng/L (n=533). No significant differences between specimen types, assay incubation time, or reagent lots existed. A substantial positive bias (76%) exists between the 4th generation and hscTnT assays at the low end of the measuring range (<50 ng/L). hscTnT serum pool concentrations were within 2SD limits of the mean of means in the comparison trial, indicating comparable results across multiple platforms and laboratories.
Conclusion:
The Roche hscTnT assay conforms to guideline precision requirements and will likely identify additional patients with myocardial injury suspicious for AMI.
More Related Videos
07:40Impact of High-intensity Interval Exercise and Moderate-Intensity Continuous Exercise on the Cardiac Troponin T Level at an Early Stage of Training
Published on: October 10, 2019
09:23Using Extraordinary Optical Transmission to Quantify Cardiac Biomarkers in Human Serum
Published on: December 13, 2017
Related Concept Videos
Acute Coronary Syndrome III: Diagnostic Studies
Blood Studies for Cardiovascular System I: Cardiac Biomarkers
The essential diagnostic tools for detecting myocardial necrosis and monitoring individuals suspected of having acute coronary syndrome (ACS) include:
Troponins
Troponins, particularly cardiac troponins I and T, are the most precise and sensitive markers of myocardial injury. They are detectable within 4-6 hours of myocardial injury and remain...