Related Experiment Video
Updated: Jun 5, 2026

Laparoscopic Non-Mesh Cerclage Pectopexy with Uterine Preservation for Pelvic Organ Prolapse
Published on: October 25, 2024
Trocarless system for mesh attachment in pelvic organ prolapse repair--1-year evaluation
Menachem Alcalay1, Michel Cosson, Miron Livneh
1Department of Obstetrics and Gynecology, Chaim Sheba Medical Center, Ramat Gan 52621, Israel. malcalay@netvision.net.il
Introduction And Hypothesis:
EndoFast Reliant system is a novel technology for pelvic organ prolapse (POP) repair that attaches mesh directly with metal fasteners.
Methods:
This was a prospective multicenter study in 20 women who underwent vaginal POP repair to evaluate safety and efficacy. The patients were followed for 12 months, using POP-Q measures, questionnaires on symptoms (PFDI), and sexual function (FSFI).
Results:
No intraoperative complications occurred. Prolapse has been anatomically resolved (< stage 2) in 85% at 1-year follow-up (17/20 patients). Pelvic Floor Distress Inventory (PFDI) scores showed significant improvement for 1 year in symptoms related to prolapse (4.1-0.7, p < 0.001) and bladder (1.4-0.7, p = 0.06) but not to rectum. Female Sexual Function Index (FSFI) scores did not change significantly. One case of mesh exposure was found (5%) and treated conservatively. No migration of fasteners was observed on X-ray.
Conclusions:
The EndoFast Reliant system was found safe and efficacious. Future studies are needed to evaluate its advantages over techniques using trocars.