Optimal gating compared to 3D and 4D PET reconstruction for characterization of lung tumours
Wouter van Elmpt1, James Hamill, Judson Jones
1Department of Radiation Oncology (MAASTRO), GROW - School for Oncology and Developmental Biology, Maastricht University Medical Centre, Dr. Tanslaan 12, NL-6229 ET Maastricht, The Netherlands. wouter.vanelmpt@maastro.nl
Purpose:
We investigated the added value of a new respiratory amplitude-based PET reconstruction method called optimal gating (OG) with the aim of providing accurate image quantification in lung cancer.
Methods:
FDG-PET imaging was performed in 26 lung cancer patients during free breathing using a 24-min list-mode acquisition on a PET/CT scanner. The data were reconstructed using three methods: standard 3D PET, respiratory-correlated 4D PET using a phase-binning algorithm, and OG. These datasets were compared in terms of the maximum SUV (SUVmax) in the primary tumour (main endpoint), noise characteristics, and volumes using thresholded regions of SUV 2.5 and 40% of the SUVmax.
Results:
SUVmax values from the 4D method (13.7 ± 5.6) and the OG method (14.1 ± 6.5) were higher (4.9 ± 4.8%, p < 0.001 and 6.9 ± 8.8%, p < 0.001, respectively) than that from the 3D method (13.1 ± 5.4). SUVmax did not differ between the 4D and OG methods (2.0 ± 8.4%, p = NS). Absolute and relative threshold volumes did not differ between methods, except for the 40% SUVmax volume in which the value from the 3D method was lower than that from the 4D method (-5.3 ± 7.1%, p = 0.007). The OG method exhibited less noise than the 4D method. Variations in volumes and SUVmax of up to 40% and 27%, respectively, of the individual gates of the 4D method were also observed.
Conclusion:
The maximum SUVs from the OG and 4D methods were comparable and significantly higher than that from the 3D method, yet the OG method was visibly less noisy than the 4D method. Based on the better quantification of the maximum and the less noisy appearance, we conclude that OG PET is a better alternative to both 3D PET, which suffers from breathing averaging, and the noisy images of a 4D PET.
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET