Validation of wearable monitors for assessing sedentary behavior
Sarah Kozey-Keadle1, Amanda Libertine, Kate Lyden
1Department of Kinesiology, University of Massachusetts, Amherst, MA, USA.
Medicine and Science in Sports and Exercise
|January 15, 2011
Summary
The activPAL monitor accurately measures sedentary behavior and detects changes in sitting time, unlike the ActiGraph GT3X. For accurate sedentary behavior assessment, the activPAL is recommended over the ActiGraph.
Area of Science:
- Physical activity monitoring
- Health behavior research
- Biomedical engineering
Background:
- Accurate measurement of sedentary behavior (SB) is crucial for understanding its health implications.
- Existing wearable monitors for assessing SB in free-living conditions often lack validated accuracy.
- This study addresses the need for reliable SB assessment tools.
Purpose of the Study:
- To validate commercially available monitors for assessing sedentary behavior.
- To compare the accuracy and sensitivity of activPAL (AP) and ActiGraph GT3X (AG) devices.
- To evaluate different cut-points for the ActiGraph device in defining sedentary behavior.
Main Methods:
- Twenty overweight, inactive office workers participated in two 6-hour observation periods.
- Participants wore both activPAL (AP) and ActiGraph GT3X (AG) monitors.
- Direct observation (DO) was used as the criterion measure, with participants instructed to reduce sitting time during the second period.
Main Results:
- The activPAL (AP) showed high agreement with direct observation (R2 = 0.94) for sedentary time.
- The ActiGraph GT3X (AG) with a 100 counts per minute cut-point underestimated sitting time and had lower agreement (R2 = 0.39).
- The AP demonstrated superior sensitivity in detecting reductions in sitting time, while the AG with 150 counts per minute showed the lowest bias among AG cut-points.
Conclusions:
- The activPAL monitor is more precise and sensitive for assessing sedentary behavior compared to the ActiGraph GT3X.
- Researchers studying sedentary behavior should consider utilizing the activPAL monitor for enhanced accuracy.
- For ActiGraph users, a cut-point of 150 counts per minute may offer a more appropriate definition of sedentary behavior.