Evaluation of intraocular lens tilt with anterior segment optical coherence tomography
Dhivya Ashok Kumar1, Amar Agarwal, Gaurav Prakash
1Dr Agarwal's Eye Hospital and Eye Research Centre, Chennai, India.
Purpose:
To determine the ability of anterior segment optical coherence tomography (OCT) to detect intraocular lens (IOL) tilt evaluation in relation to the limbus.
Design:
Observational case series.
Methods:
The IOL position of 123 eyes of 92 patients was examined with anterior segment OCT (Carl Zeiss Meditec, Dublin, California, USA). All eyes underwent uneventful phacoemulsification with the IOL in the bag. Images were obtained in 4 axes (180 to 0 degrees, 225 to 45 degrees, 315 to 135 degrees, and 270 to 90 degrees). Using MatLab software version 7.1 (Mathworks), the OCT images were analyzed. The distance between the iris margin and the anterior surface of IOL, the slope ratio between IOL and limbus, and the angle (θ; position of IOL with reference to the limbus) were determined and were correlated with the astigmatism and vision.
Results:
The mean slope of the limbus and the IOL in all axes was 0.003 ± 0.09 and -0.002 ± 0.12, respectively. The average slope ratio was 1.1 ± 1 (range, -2.09 to 3.82) and the angle (θ) was 1.52 ± 0.9 degrees (range, 0.04 to 3.6 degrees). The mean ocular residual astigmatism was 0.2379 ± 0.469 diopters. There was no significant correlation of ocular residual astigmatism with slope ratio (r = -0.171; P = .060) and slope angle (r = -0.132; P = .147). There was significant correlation of ocular residual astigmatism with total astigmatism (r=0.602, p=0.000). The mean distances between the iris margin and the anterior surface of IOL at the pupillary plane were 0.80 ± 0.6 mm and 0.83 ± 0.57 mm, respectively.
Conclusions:
The normal in-the-bag IOL maintains an angle with reference to the limbus and a slope ratio without causing a significant tilt. Anterior segment OCT can be used as an alternative in IOL tilt evaluation by the analysis of its position in relation to the limbus.