Related Experiment Video
Updated: Jun 5, 2026

Wideband Optical Detector of Ultrasound for Medical Imaging Applications
Published on: May 11, 2014
Interferometric fiber optic sensors for biomedical applications of optoacoustic imaging
Horacio Lamela1, Daniel Gallego, Rebeca Gutierrez
1Universidad Carlos III de Madrid, Madrid, Spain. horacio@ing.uc3m.es
Abstract:
We present a non-metallic interferometric silica optical fiber ultrasonic wideband sensor for optoacoustic imaging applications. The ultrasonic sensitivity of this sensor has been characterized over the frequency range from 1 to 10 MHz. A comparative analysis has been carried out between this sensor and an array of piezoelectric transducers using optoacoustic signals generated from an optical absorbent embedded in a tissue mimicking phantom. Also, a two dimensional reconstructed image of the phantom using the fiber interferometric sensor is presented and compared to the image obtained using the Laser Optoacoustic Imaging System, LOIS-64B. The feasibility of our fiber optic based sensor for wideband ultrasonic detection is demonstrated.